TECHNICAL SPECIFICATION
Specifications
IFB-51
IFB-52
Temperature range
50°C to 600°C (122°F to 1112°F)
50°C to 600°C (122°F to 1112°F)
Stability at 450°C
±1.0°C (20cm/8" immersion depth
with lid after 2 hours at set point)
±1.0°C (46cm/18" immersion depth
with lid after 2 hours at set point)
Display accuracy
±10.0°C (20cm/8" immersion depth
with lid after 2 hours at set point)
±10.0°C (46cm/18" immersion
depth with lid after 2 hours at set
Temperature d
isplay
°C
or
°F
°C
or
°F
Type of control
3 term PID
3 term PID
Sensor type
K type thermocouple
K type thermocouple
Internal diameter
255mm (10.1”)
255mm (10.1”)
Working diameter when 213mm (8.4”)
213mm (8.4”)
Depth
405mm (16”)
762mm (30”)
Working depth when
305mm (12”)
660mm (26”)
Load Capacity (1/3 of
volume)
5 litres
12 litres
Air pressure
207 to 1034 kPa (30 to 150 psi)
clean, dry and oil-free
207 to 1034 kPa (30 to 150 psi)
clean, dry and oil-free
Air Consumption at
Ambient (max)
115 litres/min
115 litres/min
Air Consumption at
48 litres/min
48 litres/min
Fluidised medium
40kg (88lb) aluminium oxide
(Alundum), 120 mesh, part code
Brown/A
L
O
73kg (160lb) aluminium oxide
(Alundum), 120 mesh, part code
Brown/A
L
O
Heat up time 20°C to
450°C
105 minutes
210 minutes
Heat up time 20°C to
600°C
150 minutes
300 minutes
Nominal heater power
4kW
6kW
Electrical supply
230V, 50/60Hz, single-phase
230V, 50/60Hz, single-phase
Exhaust fan requirements 7125 litres/min (250 CFM)
7125 litres/min (250 CFM)
Net weight
60kg (132lb)
84kg (185lb)
Gross weight incl.
aluminium oxide
100kg (220lb)
156kg (344lb)
Overall size W x D x H
521 x 521 x 686
609 x 533 x 1041