OPERATING THE BRACKET ON THE SAFETY MD TRANSPORTER

INSTALLING THE BRACKET ONTO THE SAFETY MD TRANSPORTER

Operation Guide

1. 

Only Technimount brackets with an Anti-Rotation 

Micro Disc on the back can be inserted into the 

Micro Base located on the Safety MD Transporter.

2. 

Ensure  that  your  bracket  has  an  Anti-Rotation 

Micro Disc on the back to install the medical 

device onto the Safety MD Transporter.

3. 

While  using  both  hands,  slowly  insert  Anti-

Rotation  Back  Disc  located  on  the  back  of  the 

bracket into the Micro Base.

WARNING

 

9

Ensure that the bracket is securely locked into 

the Micro Base before transport.

Anti-

Rotation 

Back 

Micro 

Disc

Summary of Contents for BRACKET PRO 101 Series

Page 1: ...BRACKET PRO SERIE 101 USER GUIDE EMS TM ...

Page 2: ...ated into any language in any form or by any means without the written permission of Technimount or its affiliate companies Trademark Attributions Bracket Pro Serie Technimount EMS are registered trademarks of Technolgies CGC Inc TV 100 is a registered trademark from Bio Med Devices In order to ensure continuous improvements to our products and services we invite you to communicate your comments t...

Page 3: ...339 6351 NOTE For any issues with your Technimount product its components or for any technical questions during the installation or maintenance processes please send an email to techsupport technimount com Please have the serial number of your Technimount product available as shown in the figure below when calling Technimount Technical Support Include the serial number in all written communication...

Page 4: ...ety MD Transporter 23 Installing the Bracket onto the Safety MD Transporter 23 Rotating the Bracket on the Safety MD Transporter 24 REmoving the Bracket with the Medical Device from the Safety MD Transporter 25 General Operating Guidelines for the Safety Arm System 26 Operating the Bracket on the Safety Arm System 27 Installing the Bracket on the Standard Surface Base 27 Rotating the Bracket on th...

Page 5: ...estic policy Warranty outside Canada may vary by country Please contact TECHNIMOUNT Customer Service for more information 36 Return Policy 37 Prior to 30 Days 37 Prior to 45 Days 37 Prior to 60 Days 37 Return Authorization 37 Damaged Merchandise 38 Claim Process 38 Required Information 38 Findings and Conclusion 38 QUESTIONS ABOUT OUR POLICY 38 SAE CERTIFICATION 39 ...

Page 6: ... standard base on a Safety MD Transporter or with Technimount s Floor Mount or Wall Mount Systems These solutions provide a higher flexibility for different EMS CCT protocols and applications The Bracket Pro Serie 101 is compatible with the floor platform of air ambulances in Europe It can also be used with the wall mount for ground ambulances in Europe and North America This mounting system is co...

Page 7: ...101 UG 202111 01 7 Return To Table of Contents Introduction POSSIBLE MOUNTING OPTIONS FOR THE BRACKET PRO SERIE 101 Safety MD Transporter Wall Mount System EU Safety Arm System Floor Mount Air Ambulance Standard Surface Base ...

Page 8: ...101 UG 202111 01 www technimount com Introduction SYMBOLS AND DEFINITIONS SYMBOL DESCRIPTION Warning and Caution special attention is required Consult accompanying documents Safe working load symbol and Load balance symbol Pinch Point ...

Page 9: ...serious injury It may also describe potential serious adverse reactions and safety hazards Caution Alerts the reader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property This includes special care necessary for the safe and effective use of the device to avoid damage that may occur as a ...

Page 10: ... Assembly is attached to the medical device according to existing installation points for mounting with 4 screws Acetal Feet to protect the disc and medical device when placed on the ground Standard Bottom Disc compatible with all of Technimount s mounting solutions New rubber pads have been added to protect the medical device from scratches ...

Page 11: ...TION CONTINUED BRACKET PRO SERIE 101 CONTINUED Back View Introduction Anti Rotation Micro Disc is used for vertical support on different Wall or Arm Systems Back Plate is installed into the main assembly with 4 Screws Bracket is made with High Density Aluminum with Silver Anodized Finish ...

Page 12: ...E J3043 26 G FAA 14 CFR 23 applicable sections Installation Device Retained on Main Assembly with Back Plate Attachment Options 2 Locking Discs Universal Bottom Disc Anti Rotation Back Micro Disc Transportation Compatibility System Compatible with Technimount s Standard Surface Base Safety Arm System For Stryker Cots Power PRO XT and MX PRO Floor Mount Platform Flight Wall Mount System Safety MD T...

Page 13: ...stomer Service customerservice technimount com Technimount E M S Holding Inc C O Regulatory Affairs 3770 Jean Marchand Street Suite 100 C Quebec QC G1J 2N3 T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com WARNING The end user and administrator are responsible to validate regulations and standards for safety in their region in order to comply with applicable safety regulations...

Page 14: ...ed properly the medical device could fall during transport and may cause injury to the patient or operator and or damage to the product 9 Do not allow untrained staff to assist in the operation of the mounting system Untrained technicians staff can cause injury to the patient or themselves 9 Do not place additional equipment or items onto the Bracket or mounts other than approved Technimount or co...

Page 15: ...emature corrosion of critical components 9 When ordering it is important to identify on which mounting system you need to mount the bracket on as there are different systems for different EMS applications e g EMS ground or air mounting systems 9 A maintenance program should be established for all Technimount equipment Preventive maintenance may need to be performed more frequently based on the lev...

Page 16: ...m must work properly before being put into service 5 Refer to the identification of components at the beginning of this user guide if needed TOOLS REQUIRED FOR INSTALLATION Below is the basic tool required for installation If you have any questions regarding the installation or need assistance please contact our Technical Support team at techsupport technimount com 9 Phillips screwdriver NOTE 9 El...

Page 17: ...al plate located on the back of the ventilator by removing all 4 screws It will not be used for this assembly Store for future use 3 Remove the power supply 4 To remove the power supply push on the lever and slide sideways to the right 5 Once these components are removed the TV 100 is ready to be installed into the bracket Installation Guide INSTALLATION OF THE BRACKET ONTO THE TV 100 VENTILATOR ...

Page 18: ...100 VENTILATOR CONTINUED 6 Place the TV 100 inside the main assembly 7 Once the TV 100 is inside the main assembly install all four Stainless steel Phillips button head screws 8 32 x 1 into the medical device 8 Screws provided are Nylon Patch screws for a better grip 9 Tighten all four screws Installation Guide ...

Page 19: ...allation Guide INSTALLATION OF THE BRACKET ONTO THE TV 100 VENTILATOR CONTINUED 10 Place back plate onto the main assembly 11 Back plate is placed properly into the main assembly 12 Install all four Stainless steel button head philips screw 10 32 x 5 8 into the main assembly ...

Page 20: ... INSTALLATION OF THE BRACKET ONTO THE TV 100 VENTILATOR CONTINUED Installation Guide 13 Tighten all four 4 screws 14 Once all screws have been installed replace the power supply back onto the TV 100 15 Slide into the openings from the side to insert back onto the TV 100 ...

Page 21: ...TV 100 VENTILATOR CONTINUED 16 The power supply is locked onto the TV 100 17 The TV 100 mounting bracket has 2 discs providing multiple configuration options Anti Rotation Micro Disc for vertical mounting systems Standard Bottom Disc for Standard Suface Base systems 18 The TV 100 and mounting system is ready for EMS transport ...

Page 22: ...ES FOR THE SAFETY MD TRANSPORTER 9 Use the Safety MD Transporter only as described in the user guide 9 Read all the labels and instructions on the product before using the Safety MD Transporter 9 Always inspect the Micro Base and the Safety MD Transporter for any issue or damage before every use 9 Installing or removing the Safety MD Transporter requires a minimum of one 1 trained operator 9 Do no...

Page 23: ...on the back can be inserted into the Micro Base located on the Safety MD Transporter 2 Ensure that your bracket has an Anti Rotation Micro Disc on the back to install the medical device onto the Safety MD Transporter 3 While using both hands slowly insert Anti Rotation Back Disc located on the back of the bracket into the Micro Base WARNING 9 Ensure that the bracket is securely locked into the Mic...

Page 24: ...RANSPORTER CONTINUED INSTALLING THE BRACKET ONTO THE SAFETY MD TRANSPORTER CONTINUED Operation Guide 4 A Bracket Pro Serie 101 installed on the Safety MD Transporter ROTATING THE BRACKET ON THE SAFETY MD TRANSPORTER 1 To rotate the medical device 360 turn the Safety MD Transporter clockwise or counter clockwise ...

Page 25: ...TER CONTINUED REMOVING THE BRACKET WITH THE MEDICAL DEVICE FROM THE SAFETY MD TRANSPORTER 1 To remove the bracket from the Safety MD Transporter press the red quick release button located on the back Micro Base 2 While holding the red quick release button lift the bracket upwards until removed and release the red quick release button ...

Page 26: ...ed while the patient is on the cot stretcher 9 Use the Safety Arm System only as described in its user guide 9 Read all the labels and instructions on the product before using the Safety Arm System 9 Always inspect the Clamp Block Safety Pin Lock Pin and Standard Surface Base before every use 9 Installing or removing the Safety Arm System requires a minimum of one 1 trained operator 9 Do not remov...

Page 27: ...e is installed Wall Mount Floor Mount Safety Arm System etc 3 To insert the bracket into the Standard Surface Base slide into the base horizontally from the front of the base 4 Ensure the bracket is locked securely in the Standard Surface Base before manipulating the device or transporting the stretcher ROTATING THE BRACKET ON THE STANDARD SURFACE BASE 1 To rotate the medical device 360 turn the b...

Page 28: ...ease Mechanism located on the front of the Standard Surface Base 2 While holding the Red Quick Release Mechanism slide outwards horizontally to remove from the Standard Suface Base 3 Hold the medical device from the middle of the unit to gently remove from the base CAUTION Do not use the top handle to remove the unit Due to the height of the medical device it might not slide easily out of the base...

Page 29: ...ailable refer to Customer Service WARNING Do not modify any components of this system Modifying the product can cause unpredictable operation resulting in injury to the patient or operator NOTE Technimount reserves the right to change Part Numbers and Products without notice Please refer to Customer Service to ensure product part number availability For any replacement parts or for repair please c...

Page 30: ...tion and ensure that the product does not stay wet longer than the cleaner manufacturer s guidelines for proper disinfecting WARNING 9 When cleaning always use appropriate personal protection equipment PPE based on established protocols e g gloves eyewear etc CAUTION 9 Do not steam clean or use ultrasonic cleaners on the system or any of its components 9 Do not immerse the metal parts components i...

Page 31: ...ervice WARNING 9 Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne pathogens and may cause injury to the patient or the operator PREVENTIVE MAINTENANCE A preventive and regular maintenance program should be established for all Technimount equipment Preventive maintenance may be required more frequently based on the usage level of the produ...

Page 32: ... at what interval to perform these inspections please contact your Technical Support Team at techsupport technimount com Check each routine and replace damaged or worn out parts if necessary Follow timetable below for maintenance either in months M or hours H in accordance with your Service s current maintenance practices and protocols ITEM ROUTINE 1 M OR 2 H 3 M OR 6 H 6 M OR 12 H 12 M OR 24 H Br...

Page 33: ...www technimount com BP101 UG 202111 01 33 Return To Table of Contents Maintenance Guide INSPECTION AND MAINTENANCE RECORD PREVENTIVE MAINTENANCE PERFORMED BY DATE TIME ...

Page 34: ...n To Table of Contents 34 BP101 UG 202111 01 www technimount com Maintenance Guide TRAINING RECORD TRAINING DATE TRAINING METHOD USER GUIDE IN SERVICE IN CLASS ETC TRAINEE NAME BASIC TRAINING TRAINING UPDATE ...

Page 35: ...purchase Some of the parts identified on the assembly drawing parts in this user guide may not be individually available for purchase Please call Technimount Customer Service 1 888 639 2758 or at customerservice technimount com for availability and pricing DESCRIPTION PART NUMBER Anti Rotation Micro Disc TCM 052 006 Standard Bottom Disc 926 00 SS Acetal Feet TCM 032 082 ...

Page 36: ...laced after the Warranty Period has expired the Warranty Period for the repair or replacement will expire three 3 months after the date of repair or replacement LIMITED RESPONSIBILITY AND WARRANTY The Limited Warranty does not apply to normal wear that should result from normal use It does not apply when the Hardware or any component is opened disassembled or repaired by someone not authorized by ...

Page 37: ...5 DAYS Product must be unused undamaged and in its original packaging The product return request must be provided in writing and it must be approved by TECHNIMOUNT prior to returning the product Returns will not be approved on modified or damaged item Charges may apply if damaged or missing items Customer is responsible for a 25 restocking fee PRIOR TO 60 DAYS Product must be unused undamaged and ...

Page 38: ...ill in the RMA form online at https www technimount com service support service support overview REQUIRED INFORMATION Serial number Invoice number Identification of the issue Upload the pictures of the issue Submit the form A confirmation of the receipt of the claim will be sent to you within 2 business days If you have not received a confirmation Email by then please contact techsupport technimou...

Page 39: ...www technimount com BP101 UG 202111 01 39 Return To Table of Contents SAE Certification ...

Page 40: ...e it matters most Technimount E M S Holding Inc 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com 2021 Technimount E M S Holding Inc All Rights Reserved Printed in Canada ...

Reviews: