18
BP122 UG 202108-01
www.technimount.com
INSTALLATION OF THE BRACKET ONTO THE BAXTER COLLEAGUE CX VOLUMETRIC
INFUSION PUMP (CONTINUED)
4.
Locate the front block and the two (2) Black
Oxide Stainless Steel Flat Head Phillips Screws
10-32 x 5/8".
5.
Align the screws and the screw holes of the main
assembly.
6. Install the front block with the two (2) screws.
NOTE
Put a few drops of medium strength thread-locking
adhesive (blue Loctite 242) on each screw before
installation to prevent the screws from loosening
during normal use.
7. Tighten the two (2) Screws.
CAUTION
9
Do not use electrical tools to install the screws.
9
Do not overtighten the screws.
8. The Baxter Colleague CX Volumetric Infusion
Pump is ready for transport once properly
inserted into its bracket and the stretcher
mounting system.
INSTALLATION GUIDE