18
SAS UG 202205-01
WARNING/CAUTION”NOTE (CONTINUED)
WARNING (CONTINUED)
9
Do not attach restraints to the base tubular frame, the Standard Surface Base, or any other place of the arm sys
-
tem. It may result in damage to the arm and/or result in injury to the patient or operator.
9
Do not place any additional equipment or objects on the arm, other than the approved Technimount device
bracket for medical devices.
9
Always reposition the arm with care. Avoid any improper pushing, violent manipulation and impacts to soroun
-
dings as it may result in an unbalanced cot/stretcher and may cause injury to the patient or operator and/or da
-
mage to the product.
9
To avoid the risk of patient injury or equipment damage, ensure that you properly mount and secure the defibril
-
lator platform (bracket) to the arm’s Standard Surface Base.
9
To avoid the risk of patient injury or equipment damage, you must make sure that the disc attached to the bracket
under the medical device is properly locked into the Standard Surface Base prior to repositioning the arm.
9
If you want to remove the arm from the system, make sure you remove the medical device from the Standard
Surface Base prior the operation, failure of removing it may result in damage to the product, or cause injury to the
operator.
9
Improper maintenance can cause injury or damage to the product. Maintain the Safety Arm System as described in
this user guide. Use only Technimount approved parts and maintenance procedures. Using unapproved parts and
procedures could cause unpredictable operation and/or injury and will void the product warranty.
9
Failure to properly clean or dispose of corrosive products will increase the risk of premature damage and may
cause injury to the patient or operator.
9
Ensure proper hand placement on a high position on the tubular frame (pole). Hands should be clear of clamp
block while loading and unloading the arm or whenever changing height position of the arm system with two or
more operators.
9
Do not adjust or change the position of the arm while you are walking or working with the cot/stretcher. Stay with
the patient and control the arm at all times.
SIDE VIEW OF THE CLAMP BLOCK AND ARM
PIVOT
SIDE VIEW OF THE CLAMP BLOCK AND ARM
PIVOT
Summary of Safety Precautions