TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
8
The expiratory phase can be modifying adding Positive Pressure at the End of Expiration
(PEEP). The NEUMOVENT Graph Ventilator produces this positive pressure by means of a
digital regulation of the closing force of the expiratory valve diaphragm.
Pressure curves during ventilation with
5 cm H
2
O of PEEP.