TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
27
Items to be checked during maintenance
every 5000 hours of use:
1-External check-up, clean-up and disinfection of the device.
2-Functional control.
3- Control of alarms sound:
Press
“
Ctrl
”
+
“
Reset
”
4-Functional inspection of keys.
5-Leds inspections:
a-Go to Analog Input in Calibration Menu ( see chapter 4°)
b-Press
“
Silence
”
key , and check all leds (except :Technical failure,
battery and charging leds)
6-Check-up the sensors following chapter 4°.
7-Check up for software updates
.
8-Removal of back cover and clean-up the fan and area
.
9-Inspection and clean-up porous metal filters and water trap
.
10-Visual inspection of internal tubing:
Look for water and oil traces
16-Replacement of sealing joint of cabinet.
17-Reemsambl the ventilator.
18-Complete recalibration.
19-Replacement of expiratory valve.
20-Update of records and documentation.
21-Final control.
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...