TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
38
Sensor Inspection
Proportional valves:
AirVal:
1)-
Press the
“
Reset
”
key.
At that moment, the
Calibration
Menu
is displayed on the screen.
2)-
Choose the option
“Air Val”
with the
“
selection
”
keys
3)-
Put target value in 5 L/m and press the
“
fiO
2
”
key.
4)
- The valve must be open.
5)-
Press the
“
Reset
”
key to finish and press one more time to go back to the calibration menu.
OxyVal:
Idem AirVal