TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
52
How to Open and Close the
Neumovent Graph
Ventilator
Operation
Description
100
Removal of Shield- Front Cabinet
1- Remove the washer screws and foam filler.
2- Do not remove the hexagonal fillers since they do not hold the panel.
3- Remove the nuts which hold the flanges, and finally remove the flanges and white head
screws.
- The panel can be now removed. Proceed with extreme care since it is attached to the cabinet
with silicone sealer.
Flange
Flange
Fig.:10-A