TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
70
Measurement, Inspection and Test Equipment
1-Flow analyzer (TSI Certifier)
2-Oximeter
3- Medical grade Air
4-Oxigen
5-Reusable adult respiratory circuit
6-O
2
cell
7-5 Kg./cm
2
Pressure regulator
8-100 cm. H
2
O Pressure regulator
Remarks
:
- The EMI must be periodically calibrated by an official organization and the calibration
must be recorded and filed.
-The air used must be free of oil, humidity, and particles larger than 0.3 m
Procedure for Initial Calibration and Setup
Neumovent Graph Ventilator
Operación
Descripción
10
Access to the Calibration and Setup Menu:
When the device is off, press the
“
Reset
”
key, and turn it on while keeping the key pressed.
The next message will be displayed on the screen:
Pasword:
1-
If the changes to be made are to be saved in the memory, press the following sequence
of keys:
I : E
–
FREEZE
–
VOLUME - VERT
The next message will be displayed on the screen:
Correct Pasword
PRESS ANY KEY
You will get to
MENU CALIBRATION
by pressing any key
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...