TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
74
Procedure for Initial Calibration and Setup
Neumovent Graph Ventilator
Operation
Description
50
Air Flow
a-
Connect the device outlet to a standard flowmeter
Fig.:1-A (end of chapter ).
b-
Connect the inlets of the gas supply to a compressed air supply source
c-
Select the target value with
“
Selection
”
key and press
“
Enter
”
Open and close the valve with
“
Selection
”
key
Low
Fast
d-
When the target value is
taked press “
Enter
”.
e-
Select a new target and repeat the operations
a-b-c
.
f-
Press
“
Menú
”
and the next message will be displayed on the screen:
SAVE
EXIT
Select the desired option with the
“
Selection
” key
and press
“
Enter
”.
-If you select
EXIT
and
the values was modified; the next message will be displayed on
the screen:
¿Exit without save changes?
[ Enter ]
Accept
[ Reset ]
Cancel
Select the desired option
Then, press
“
Reset
”
to return to initial menu.
Air Flow
Target
Apertura
0
0
0
0
Memory
Flow
Lectura
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...