TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
75
Procedure for Initial Calibration and Setup
Neumovent Graph Ventilator
Operation
Description
60
Air Valve
a-If the valve
don’t have previous calibration
the next message will be displayed
on the screen:
START
SAVE
EXIT
b-If the valve have previous calibration then press
“
Menú
”.
c-Select
START
and press
“
Enter
”
to auto calibration begins
d-Press
“
Menú
”
and the next message will be displayed on the screen:
SAVE
EXIT
Select the desired option with the
“
Selection
” key
and press
“
Enter
”.
-If you select
EXIT
and
the values was modified; the next message will be displayed on
the screen:
¿Exit without save changes?
[ Enter ]
Accept
[ Reset ]
Cancel
Select the desired option
Then, press
“
Reset
”
to return to initial menu.
Air valve
Target
Apertura
0.0
0.1
0
0
Memoria
Flujo
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...