TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
84
Final Control Neumovent Graph Ventilator
Operation
Description
150
MMV Mode
Programming: VE = 7 L / min.
Min. VE Alarm = minimum value
PSV = 10 cm H2O
<ENTER>.
Trigger the test lung at a low frequency to obtain a VE lower than that programmed.
Ensure:
a) Peak pressure should increase breath by breath.
b) VT should increase breath by breath.
160
PSV + VT Assured mode
Programming:
Accept default parameters (CTRL + ENTER).
Trigger the test lung.
Control activation of the continuous flow phase.
Check: VT expected = VT target
170
Check operation of serial output.
180
Control the operation of the device installed with a circuit and test lung for 48 hrs.