and physiological, forming metaemoglobina which is no longer able to carry oxygen to the tissues. Also, low con-
centrations (4 ppm) will anesthetize the nose. 

OPERATIONAL DESCRIPTION 

The electrochemical sensor is temperature compensated, but is sensitive to extreme humidity variations. The cali-
bration is carried out  with specific gas to be detected. Anyway, it can contemporaneously detect other gases that 
should be present in the same environment as listed in

 

tables 3 and 4

Note

 that the Led are not visible when the enclosure is closed. 

Preheating

: when powered, the sensor needs a time of preliminary heating of about 60 seconds. During this peri-

od the yellow LED “

FAULT

” flashes. After this period, the yellow LED light off, the green LED “

ON

” illuminates to in-

dicate normal functioning. After this period the unit is able to detect gas even if it attains the optimum stability con-
ditions after about 4 hours continual functioning.  

Normal operation

: the green LED “ON” should be light on. 

ALARMS:

 different alarms levels, depending on the models, can be set by Dip-Switch (

see tables 2 and 5÷11

). There 

are two different alarm methods, the normal one, will activate the corresponding relays after a fixed delay, the ad-
vanced one, operate with TLV-TWA, STEL and C value, in this case the relays will activate without delay. 

The 1st  red Led

 (

ALARM 1

) illuminates  when the  gas  concentration attains 1st  alarm level. Under  normal alarm 

method, after about 12 seconds the “

ALARM 1

" relay will activate, while under advanced one, with TLV-TWA level 

selected, the relay will activate without delay. 

The  2nd  red  Led

  (

ALARM  2

)  illuminates  when  the  gas  concentration  attains  2nd  alarm  level  and  after  about  30 

seconds, the “

ALARM 2

" relay will activate or without delay when TLV-STEL level is selected. 

The  3rd  red  Led

  (

ALARM  3

)  illuminates  when  the  gas  concentration  attains  3rd  alarm  level  and  after  about  60 

seconds, the “

ALARM 3

" relay will activate or without delay when TLV-C level is selected. 

TLV (

T

hreshold 

L

imit 

V

alues

)

 are defined as an exposure limit to which it is believed nearly all workers can be exposed 

day after day for a working lifetime without ill effect. 

TLV-TWA (

T

hreshold 

L

imit 

V

alue–

T

ime-

W

eighted 

A

verage

)

 is the time-weighted average concentration for a conventional 

8-hour workday and a 40-hour workweek, to which it is believed that nearly all workers may be repeatedly exposed, 
day after day, without adverse effect. 

TLV-STEL

 

(

T

hreshold 

L

imit 

V

alue–

S

hort-

T

erm 

E

xposure Limit

)

 is the concentration to which it is believed that workers can 

be  exposed  continuously  for  a  short  period  of  time  without  suffering  from  irritation,  chronic  or  irreversible  tissue 
damage, or narcosis. STEL is defined as a 15-minute TWA exposure, which should not be exceeded at any time 
during a workday. 

TLV-C (

T

hreshold 

L

imit 

V

alue-

C

eiling

)

 is the concentration that should not be exceeded during any part of the working exposure. 

The  values  are  recommending  exposure  levels  that  are  protective  to  workers,  OSHA  (

O

ccupational 

S

afety  and 

H

ealth 

A

dministration

, of the U.S. Department of Labour) and COSHH 

(

C

ontrol Of 

S

ubstances 

H

azardous to 

H

ealth in Europe

).

 

Faults

the  instrument  signal  different  kind  of  failures,  as  listed  below.  The  yellow  Led  illuminates,  the  "

S

"  output 

falls down to 0mA and the “

FAULT

 normally activated relay deactivate. This relay, if necessary, can be used both to 

signal remotely an occurred damage and to signal the absence of power to the instrument. 

Yellow  Led  illuminates  each  4  seconds  (

with  green  Led  activate

):

 

this  happens  when  the  “

Cartridge  Sensor

”  has 

overcome its theoretical period of life and its correct operation is not longer guaranteed. The detector keeps on op-
erating normally but it is necessary to replace, as soon as possible, the “

Cartridge

” with a new one. The type to be 

required is described on 

Page 1

. The replacement procedure is described in the attached manual. 

Yellow Led activate, green Led off (

FAULT relay activates and 0 mA output signal

):

 

this signal different kind of faults. 

1)

 The Dip-Switch set up is wrong, please verify (

see Tables 2 and 5÷11

). 

2)

 The “

Cartridge

” is not working, please re-

place  with  new  one. 

3)

 

If  a  new

 

Cartridge

”  is  installed  or  it  is  not  correctly  connected  or  a  not  compatible  one  is 

mounted. Please check the cartridge connections and compatibility (

see on page 1

) these checks are made switching 

off and on the device.

 

If the condition does not change, it will be necessary to replace the unit and/or send it back 

to the supplier to repair. 

Yellow and green Led activates (

FAULT relay activates and 0 mA output signal

):

 

this happens when the “

Cartridge

” is 

not working. First try to perform the procedure of “

ZERO

” as described in the section

 “

Test and Calibration

 

> Zero ad-

just

 

then  disconnect  and  connect  the  unit,  finally  try  to  replace  a  new 

Cartridge

”.  If  the  condition  is  not  change, 

please replace

 

the unit and/or send it back to the supplier to repair. 

All Led activates (

FAULT relay activates and >20 mA output signal

):

 

this happens when the “

Cartridge

” is not working or 

gas concentration is higher than F.S. If there are not any gas leaks and the condition is not change, it will be nec-
essary to replace the unit and/or send it back to the supplier to repair. 

INSTALLATION 

The detector must be accurately installed and testing according to the national dispositions in force on the safety of 
the plants and installation of electric devices in areas with danger of explosion. 

Mounting

: The 

Fig. 2

 shows the instrument size. The unit must be positioned vertically with the sensor downwards.  

SE138EA and SE138EA-H positioning: 

should be fixed at 30÷50 cm from the ceiling (

NH

3

 gas is lighter than air

). 

SE138EC-S and SE138EC-H positioning: 

should be fixed at 1.60 m from the floor (

CO

 

gas is a little lighter than air). 

SE138ES positioning: 

should be fixed at 30÷50 cm from the floor (

SO

2

 gas is heavy as air) 

SE138EH positioning: 

should be fixed at 30-50 cm from the floor (

H

2

S

 gas is heavier than air). 

SE138EHCN positioning: 

should be fixed at 1.60 m from the floor for the protection of persons or over the likely 

point of emission. (

HCN 

gas is a little lighter than air

). 

SE138EN positioning: 

should be fixed at 1.60 m from the floor 

(

NO

 gas is heavy as air). 

SE138EN2 positioning: 

should be fixed at 

30÷50

 cm from the floor 

(

NO

2

 gas is heavier as air). 

Summary of Contents for SE138E Series

Page 1: ...ostituibile Inside Replaceable Cartridge Sensor Avec Cartouche Capteur échangeable Sommario Contents Sommaire Caratteristiche tecniche Technical specifications Caractéristiques techniques 2 IT DESCRIZIONE 3 NOTE SUL GAS RILEVATO DAI VARI MODELLI 3 FUNZIONAMENTO 4 INSTALLAZIONE 4 AVVERTENZE 5 VERIFICHE E CALIBRAZIONE 5 EN DESCRIPTION 7 NOTES ON THE DETECTED GAS 7 OPERATIONAL DESCRIPTION 8 INSTALLAT...

Page 2: ... pulita Average Life in fresh air Vie moyenne en air pur 2 anni years ans 2 anni years ans 3 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi month mois 1 mese month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mes...

Page 3: ...iente è respirato dalle persone dai polmoni raggiunge il sangue dove impedisce il trasporto dell ossigeno ai tessuti e agli organi SE138EC H è identico al precedente ma utilizza un sensore di livello superiore adatto a industrie e ambienti con presenza d inquinati e altri gas tossici SE138ES ha un sensore in grado di rilevare Biossido di zolfo SO2 sinonimi Anidride solforosa Ossido di zolfo Ossido...

Page 4: ... subire irritazioni danni cronici irreversibili o narcosi TLV C Ceiling è il Limite massimo di concentrazione che non deve mai essere superata I valori indicati sono riferiti alle prescrizioni degli enti che si occupano della salute dei lavoratori lo statunitense OSHA Occupational Safety and Health Administration e l europeo COSHH Control Of Substances Hazardous to Health I Guasti possibili sotto ...

Page 5: ...EA SE138EA H SE138EC S SE138EC H SE138ES SE138EH SE138EHCN SE138EN SE138EN2 500 ppm NH3 1000 ppm CO 150 ppm SO2 500 ppm H2S 100 ppm EHN 1000 ppm NO 150 ppm NO2 Attenzione i sensori elettrochimici contengono una piccola quantità d acido Nel caso avvenga una perdita dal sensore causata da un azione meccanica violenta o dall uso oltre i limiti di funzionamento evitare di toccare il li quido fuoriusci...

Page 6: ...dere circa 3 5 minuti fino quando il 3 Led rosso si accende fisso e mentre è ACCESO premere il tasto F2 tenendolo premuto finché 1 e 3 Led rossi non rimangono spenti per almeno 2 secondi se invece il 1 Led rosso continua a lampeggiare oltre 5 6 minuti significa che il flusso di gas non è costante o insufficiente o la concentrazione non è quella richiesta Interrompere l operazione spegnere lo strum...

Page 7: ...ur dioxide SO2 also Sulphurous anhydride Sulphurous oxide Sulphur ox ide gas which is a toxic colourless with pungent odour heavier than air its density as to air is 2 25 His TWA Time Weighted Average is 2 ppm and STEL Short Term Exposure Limit is 5 ppm SO2 is a strong irritant to eyes and respiratory system Inhalation may cause pulmonary edema and prolonged ex posure can lead to death and when it...

Page 8: ...hat are protective to workers OSHA Occupational Safety and Health Administration of the U S Department of Labour and COSHH Control Of Substances Hazardous to Health in Europe Faults the instrument signal different kind of failures as listed below The yellow Led illuminates the S output falls down to 0mA and the FAULT normally activated relay deactivate This relay if necessary can be used both to s...

Page 9: ...Cell contains a small quantity of acid It is possible that either a very strong mechanical action or an incorrect instrument use beyond the limits of its operational functioning could provoke a liquid leak out of the sensor Should this event occurs please avoid touching the come out liquid Should a contact with either the skin or eyes occur wash immediately and abundantly with water Cross sensitiv...

Page 10: ...st and 3 rd red Led are switched off for at least 2 seconds but if the 1 st red Led continues to blink more than 5 6 minutes means that the gas flow is not constant or unsatisfactory or the concentration is not as required stop the operation turn off the instrument and recalibrate Then close the gas cylinder and remove TC011 At this point we can have two possibilities Yellow and green Led illumina...

Page 11: ... mais à des concentrations élevées en effet sa LIE Limite Inférieure d Explosivité est 10 9 volume Sa Limites d Exposition Professionnelle LEP est TWA Time Weighted Average 25ppm et STEL Short Term Exposure Limit 50ppm Le CO est produit par la combustion incomplète des combustibles à base de carbone Quand ces combustibles gaz huile bois et charbon ne brûlent pas correc tement il se forme un excès ...

Page 12: ...n de gaz dépasse le second seuil si la concentration se maintient environ 30 secondes le 2ème relais ALARM 2 intervient En outre avec la méthode avancée il intervient sans retard quand le niveau TLV STEL est dépassé Le 3ème led rouge ALARM 3 s allume si la concentration de gaz dépasse le troisième seuil si la concentration se maintient environ 60 secondes le 3ème relais ALARM 3 intervient En outre...

Page 13: ...lise le détec teur avec les Dip Switch dans une position réservée non permise l indication de Dérangement sera acti vée voir Fonctionnement Dérangement Important Une fois terminée l installation alimenter le détecteur attendre 30 60 minutes environ et puis pour adapter le capteur aux conditions ambiantes exécuter la Régulation du Zéro Vérifications et Calibrage INSTRUCTIONS La vie utile du capteur...

Page 14: ...t chaque semestre en fonction des conditions ambiantes Considérer que les SE138E sont dotée d un suiveur de zéro sophistiqué qui toutes les heures refait le zéro du capteur si naturellement il n y a pas de présence de gaz Remarque Il n est pas possible d exécuter l opération si la 1ère et ou la 2ème et ou la 3ème led rouges sont déjà allu mées ou si la sortie en mA est supérieure a la valeur indiq...

Page 15: ...10 mV 19 21 mA 300 ppm 20 SE138EC H SE138ES 10ppm Dioxyde de soufre SO2 dans azote 0 3 l min 118 122 mV 11 7 12 3 mA 10 0ppm 0 2 SE138EH 50 ppm Sulfure d hydrogène H2S dans azote 0 3 l min 117 123 mV 11 7 12 3 mA 50 ppm 2 SE138EHCN 10 ppm Cyanure d hydrogène HCN dans azote 0 3 l min 192 208 mV 19 2 20 8 mA 10 ppm 0 5 SE138EN 100 ppm l Oxyde nitrique NO dans azote 0 3 l min 192 208 mV 19 2 20 8 mA ...

Page 16: ...m 4 20 10 1 5 5 Cloro Clorine Chlore 1 ppm 1 Etilene Ethylene Éthylène 100 ppm 0 1 25 0 Idrogeno Hydrogen Hydrogène 1000 ppm 1 0 6 Idrogeno Solforato Hydrogen sulphide Sulfure d hydrogène 15 ppm 15 90 1 5 19 Monossido di carbonio Carbon Monoxide Monoxyde de carbone 300 ppm 3 15 0 0 8 Ossido di azoto Nitric oxide Oxyde nitrique 35 ppm 0 7 28 0 35 0 Toluene Toluene Toluène 50 ppm 1 5 Tabella 5 Table...

Page 17: ... ON TLV TWA OSHA 10 ppm TLV STEL OSHA 15 ppm TLV C 50 ppm 0 0 0 30 OFF ON ON Risevato Reserved Réservé Risevato Reserved Réservé ON ON ON Risevato Reserved Réservé Risevato Reserved Réservé Tabella 9 Table 9 Tableau 9 SE138EHCN Acido cianidrico Hydrogen cyanide Cyanure d hydrogène S1 SET Dip Switch Livelli Allarme Relé Alarm levels Relays Niveaux d alarmes Relais Ritardo dei Relé in Secondi Relays...

Page 18: ...é COSHH Ente Europeo European Department Organisme européen OSHA Ente Americano U S Department Organisme Etats unien 190 65 83 105 70 Ø42 107 Fig 1 Dimensioni Size Dimensions Imbocco Femmina Gas Conico UNI 6125 UNI 6125 Female Gas Conical Thread Entrée femelle Gaz Conique UNI 6125 S 12 24V mA 1 2 3 4 ON OFF SET Dip Switch TP1 TP2 Fig 2 Schema di collegamento Wiring diagram Schéma AL2 AL3 AL1 FAULT...

Reviews: