MODEL 

Sample gas bottle 

Tecnocontrol Code 

TS220EA - TS220EA-H 

100 ppm  Ammonia (NH

3

) in 

Nitrogen 

Available on request

 

TS220EC-S - TS220EC-H 

300 ppm  Carbon monoxide (CO

)

 

in Nitrogen 

BO210 

TS220ES 

10 ppm 

Sulphur dioxide (SO

2

in Nitrogen 

Available on request

 

TS220EH 

50 ppm 

Hydrogen sulphide (H

2

S) in Nitrogen 

Available on request 

TS220EHCN 

10 ppm 

Hydrogen cyanide (HCN)  in Nitrogen 

Available on request 

TS220EN 

100 ppm  Nitric oxide (NO) 

in Nitrogen 

Available on request 

TS220EN2 

30ppm 

Nitrogen dioxide (NO

2

) in 

air 

Available on request 

TS220EHCl 

10ppm 

Hydrogen cloride (HCl) 

in Nitrogen 

Available on request 

It is possible to use either the 

disposable cylinders with adjust valve

 or the high pressure ones with reduction gear. 

Is also necessary the 

Tecnocontrol calibration kit model TC011

. To ensure that the gas concentration reaches 

sensor, use steel reduction gear and Teflon tubes or other compatible materials which do not disperse the gas. 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the detector. After having inserted the "

Check C

ode

", all Led are switched off. Then Led will switch on in se-

quence, from the yellow up to the red. At the end all the Led will remain lighted for around 5 seconds, then the de-
tector returns at the conditions of normal operation. It is advisable to perform this operation every 12 months ac-
cording to the use. 

Note

: this function is not working if the red LED is already switched on. 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

this function is to adjust the Zero sensor and can be done in clean 

air only (

environment without the presence of toxic gas or other pollutants

). Consider that the TS220E series have a sophisti-

cated follower of zero, that every hour will reset the sensor, if of course there is the presence of gas. If necessary, 
perform this operation after installation or after changing the “

Cartridge

” or every 6-12 months depending on envi-

ronmental conditions. After entering the "

Zero Code

" as confirmation of the operation, there will be a flashing red Led 

and output becomes 4.0 mA. 

Note

: This function is not working if mA output is higher than the value shown in the table below, depending on the 

model, or if the red Led is already ON. In this case, please recalibrate the unit and/or replacing the “

Cartridge

” 

TS220EA 

TS220EA-H 

TS220EC-S 

TS220EC-H 

TS220ES 

TS220EH 

TS220EHCN 

TS220EN 

TS220EN2 

TS220EHCl 

4,8 mA 

(15 ppm NH

3

4,8 mA 

(15 ppm CO) 

6,4 mA 

(3 ppm SO

2

)

 

4,8 mA 

(5 ppm H

2

S)

 

8 mA 

(2,5 ppm HCN)

 

5,6 mA 

(10 ppm NO)

 

6,6 mA 

(5 ppm NO

2

)

 

8 mA 

(2,5 ppm HCl)

 

 “CALIBRATION” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the sensor 

using the gas mixture indicated above.

 

Warning

to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-

low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and 
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation. 

Important note:

 

during Calibration routine the mA output indicates 0mA

.

 

The “

Calibration

” can be done in clean air only (

environment without the presence of toxic or other polluting gas

). With the keys 

perform the “

Calibration Code

". Wait until the yellow and green Led switch on fix and the red Led starts to flash. In-

sert the TC011 on the sensor holder and adjust the influx of the gas the way that the flow meter reaches the value 
indicates in table below in 

column 3

, depending on the model (

see also Fig.3

). 

Wait around 3÷5 minutes

, until the 

red Led lights (

and while it is ON

), press the key F2 and hold it pressed until the red LED is switched off for at least 2 

seconds (

but if the red LED continues to blink more than 5-6 minutes means that the gas flow is not constant or 

unsatisfactory, or the concentration is not as required, stop the operation, turn off the instrument and recalibrate

). 

Then, close the gas cylinder and remove TC011

. At this point we can have two possibilities: 

Yellow and green Led illuminates:

 the calibration routine is correct. 

Wait 8 seconds, until the instrument auto-

matically restores the normal working conditions

(

See “

Operational Description > Preheating

). 

Yellow Led illuminates:

 the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-

peat 

Preheating

, then repeat the “

Calibration

 routine without inserting again the code. If condition still persists after 

the replacement of the “

Cartridge

”, please send the detector back to the manufacturer for reparation. 

“CALIBRATION CHECKS” (

no Code required

):

 

using the gas mixture indicated above, this function is used to con-

trol, the correct response to the gas and can be made after the "

Calibration

" or the installation, but should be done 

during the periodic maintenances, as it is the only method to verify the effective functioning of the detector.

 

Insert the TC011 on the sensor holder and adjust the sample gas bottle’s valve, so that the flow meter reaches the 

value indicates in table below in 

column 3

, depending on the model (

see also Fig.3

). Check that the voltmeter, with 

probes connected on tests points “

TEST mV

”, 

(

see fig.2

reaches (in 2÷3 minutes) the value shown in 

column 4

, de-

pending on the model, [

corresponding to 

mA

 output indicated in 

column 5

 and the Central Unit which is connected to the detector, dis-

plays the value indicated in 

column 6

]. If the result is different, is necessary to recalibrate the sensor (

see “

Calibration

). 

Then, close the gas bottle, remove TC011. The mA output will slowly decrease up to 4 mA (

normal working conditions

). 

1-Model 

2-Sample gas bottle to be use 

3-Flowmeter ADJ  4-TEST mA  5-mA Output  6-Central Unit Display 

TS220EA 

100ppm Ammonia (NH

3

) ) in Nitrogen 

0.3 l/min 

90÷96 mV 

9.0÷9.6 mA 

100 ppm 

(±5)

 

TS220EA-H 

1 l/min 

TS220EC-S 

300ppm Carbon monoxide (CO) in Nitrogen 

0.3 l/min 

190÷210 mV 

19÷21 mA 

300 ppm 

(±20)

 

TS220EC-H 

Summary of Contents for TS220E

Page 1: ...on altri Gas T leau 4 r s Manual 20mA di ga Gas Detecto metteurs 4 RATE MI it e m e Attentame Please re e avec soin e alibrato per alibrated for Ammoniaca A Ammoniaca A onossido di ca onossido di ca Biossido di zolf rogeno Solfor Acido cianidric ssido di azoto iossido di azo Acido cloridrico Con Cartu e Cartridge S ommaire pecifications MODELLI DU GAZ DÈTEC Table 2 Typica Table 3 Typica Manuel d u...

Page 2: ...1000 ppm 1000 ppm 150 ppm 500 ppm Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 2 anni years ans 2 anni years ans 3 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi month mois 1 mese month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois Vari...

Page 3: ...érive à long terme 2 mese month mois 2 mese month mois 2 mese month mois 3 mese month mois Risoluzione Resolution Résolution 0 1 ppm 1 ppm 1 ppm 0 7 ppm Tempo di risposta T90 Response Time T90 Temps de réponse T90 120 secondi Seconds secondes 60 secondi Seconds secondes 60 secondi Seconds secondes 2 70 secondi Seconds secondes Ripetibilità Repeatability Répétitivité 2 del segnale signal 2 del segn...

Page 4: ...gli occhi e le vie respiratorie Per inalazione può causare edema polmonare e una prolungata esposizione può portare alla morte Nel caso si combini con l acqua forma Idrogeno Solforato H2S TS220EH ha un sensore in grado di rilevare Idrogeno Solforato H2S sinonimo Acido Solfidrico che è un gas tossi co e infiammabile incolore con un odore caratteristico di uova marce più pesante dell aria ha densità...

Page 5: ... come descritto nella sezione Verifiche e Calibrazione poi spe gnere e riaccendere l apparecchio e infine provare a sostituire la Cartuccia Sensore Se la condizione persiste sarà necessario sostituire e o inviare il rilevatore al fornitore per la riparazione Se è stato abilitato il Led rosso e quindi tutti i Led sono accesi uscita 20mA indica o guasto della Cartuccia Sensore o una concentrazione d...

Page 6: ...are in funzio ne del modello installato è MODELLO Bombola di Gas Titolato Codice Bombola TS220EA TS220EA H 100 ppm Ammoniaca NH3 in azoto Fornibile a richiesta TS220EC S TS220EC H 300 ppm Monossido di carbonio CO in azoto BO210 TS220ES 10 ppm Biossido di zolfo SO2 in azoto Fornibile a richiesta TS220EH 50 ppm Idrogeno solforato H2S in azoto Fornibile a richiesta TS220EHCN 10 ppm Acido cianidrico H...

Page 7: ...azoto 0 3 l min 90 96 mV 9 0 9 6 mA 100 ppm 5 TS220EA H 1 l min TS220EC S 300ppm Monossido di carbonio CO in azoto 0 3 l min 190 210 mV 19 21 mA 300 ppm 20 TS220EC H TS220ES 10ppm Biossido di zolfo SO2 in azoto 0 3 l min 118 122 mV 11 7 12 3 mA 10 0 ppm 0 5 TS220EH 50ppm Idrogeno solforato H2S in azoto 0 3 l min 117 123 mV 11 7 12 3 mA 50 ppm 2 TS220EHCN 10ppm Acido cianidrico HCN in azoto 0 3 0 5...

Page 8: ...ppm NO2 combined with water turns into nitric acid and nitric oxide NO2 is formed in most combustion processes using air as the oxidant At elevated temperatures nitrogen combines with oxygen to form nitrogen dioxide The most important sources of NO2 are internal combustion engines power plants heating and vehicles especially diesel engines NO2 is toxic by inhalation it can cause irreversible damag...

Page 9: ...hm with 24Vdc power supply We suggest using 24Vdc power supply WARNING Average life In clean air and in normal working condition the sensor s life is about 2 years After this period the yellow Led FAULT flashes every 4 seconds is necessary replacing the Cartridge Sensor Periodical testing we advise to carry out working tests every 6 or 12 months Tests Zero Adjust and Calibration with sample gas bo...

Page 10: ...t during the Calibration the yel low Led switch off every 8 seconds in this case interrupt the procedure switch off and switch on the instrument and repeat the Calibration If condition persists it will be necessary to send the detector to the supplier for the reparation Important note during Calibration routine the mA output indicates 0mA The Calibration can be done in clean air only environment w...

Page 11: ...fet sa LIE Limite Inférieure d Explosivité est 10 9 volume Sa Limites d Exposition Professionnelle LEP est TWA Time Weighted Average 25ppm et STEL Short Term Exposure Limit 50ppm Le CO est produit par la combustion incomplète des combustibles à base de carbone Quand ces combustibles gaz huile bois et charbon ne brûlent pas correc tement il se forme un excès de CO Quand le CO en les répandant dans ...

Page 12: ...on de gaz dépasse la valeur indiquée dans le tableau ci dessous selon le modèle Abilitation avec Dip Switch n 1 ON TS220EA TS220EA H TS220EC S TS220EC H TS220ES TS220EH TS220EHCN TS220EN TS220EN2 TS220EHCl 50 ppm NH3 200 ppm CO 10 ppm SO2 20 ppm H2S 5 ppm HCN 25 ppm NO 30 ppm NO2 5 ppm HCl Dérangement La centrale signale les anomalies décrites ci dessous en allumant la led jaune FAULT en por tant ...

Page 13: ... H TS220EC S TS220EC H TS220ES TS220EH TS220EHCN TS220EN TS220EN2 TS220EHCl 500 ppm NH3 1000 ppm CO 150 ppm SO2 500 ppm H2S 100 ppm EHN 1000 ppm NO 150 ppm NO2 30 ppm HCl Attention Les capteurs électrochimiques contiennent une petite quantité d acide sulfurique qui est corrosif Nor malement il ne doit y avoir aucune perte au capteur dans le cas contraire ceci est du à une action mécanique vio lent...

Page 14: ... dessous dans la colonne 3 voir aussi Fig 3 Attendre 3 5 minutes envi ron puis quand la led rouge s allume et pendant qu elle est allumée appuyer sur la touche F2 et maintenir cette pression tant que la led rouge ne reste pas éteinte durant au moins 2 secondes si en revanche la led rouge conti nue à clignoter plus de 5 6 minutes cela signifie que le flux de gaz n est pas constant ou insuffisant ou...

Page 15: ... Cloro Clorine Chlore 1 ppm 0 0 0 Etanolo Ethanol Éthanol 200 ppm 0 0 2 5 0 Etilene Ethylene Éthylène 100 ppm 0 0 80 50 Idrogeno Hydrogen Hydrogène 100 ppm 0 0 2 46 40 Idrogeno Solforato Hydrogen sulphide Sulfure d hydrogène 10 ppm 2 0 2 0 0 5 0 3 4 Monossido di carbonio Carbon Monoxide Monoxyde de carbone 100 ppm 0 1 100 100 1 Ossido di azoto Nitric oxide Oxyde nitrique 35 ppm 0 0 35 3 0 Tabella ...

Page 16: ...zione gas Gas Central Unit Centrale de détection RL S S Uscita 4 20 mA 4 20 mA Output Sortie en 4 20 mA Dip Switch mV Fig 3 Tester di calibrazione Calibration Tester Kit de Calibration F1 F2 Tasti Key Touches 0 3 L min 1 L min Alimentazione Power Supply Alimentation 100 300 400 500 50 0 5 200 Vdc RL Resistenza di carico Load resistor Resistance de charge RL Ohm Fig 4 Alimentazione Resistenza di Ca...

Reviews: