Calibration Kit, Sample Gas Bottles

 (

for Calibration Check and Calibration

) the mixture to be used depending  on the 

model installed is shown in the table below: 

MODEL 

Sample gas bottle 

Tecnocontrol Code 

TS293EA - TS293EA-H 

100 ppm 

Ammoniaca (NH

3

)

 

in 

Nitrogen 

Available on request

 

TS293EC-S - TS293EC-H 

300 ppm 

Carbon monoxide (CO

)

 

in 

Nitrogen 

BO210 

TS293ES 

10 ppm 

Sulphur dioxide (SO

2

in 

Nitrogen 

Available on request

 

TS293EH 

50 ppm 

Hydrogen sulphide (H

2

S)  in Nitrogen 

Available on request 

TS293EHCN 

10 ppm 

Hydrogen cyanide (HCN)  in Nitrogen 

Available on request 

TS293EN 

100 ppm  Nitric oxide (NO) 

in Nitrogen 

Available on request 

TS293EN2 

30ppm 

Nitrogen dioxide (NO

2

in air 

Available on request 

It is possible to use either the 

disposable cylinders with adjust valve

 or the high pressure ones with reduction gear. 

Is  also  necessary  the 

Tecnocontrol  calibration  kit  model  TC011

.  To  ensure  that  the  gas  concentration  reaches 

sensor, use steel reduction gear and Teflon tubes or other compatible materials which do not disperse the gas. 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of  the  detector.  After  having  inserted  the  "

Check  C

ode

",  all  Led  are  switched  off.  Then  Led  will  switch  on  in  se-

quence, from the yellow up to the red. At the end all the Led will remain lighted for around 5 seconds, then the de-
tector  returns  at  the  conditions  of  normal  operation.  It  is  advisable  to  perform  this  operation  every  12  months  ac-
cording to the use. 

Note

: this function is not working if the red LED is already switched on. 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

this function is to adjust the Zero sensor and can be done in clean 

air only (

environment without the presence of toxic gas or other pollutants

). Consider that the TS293E series have a sophisti-

cated follower of zero, that every hour will reset the sensor, if of course there is the presence of gas. If necessary, 
perform  this  operation  after  installation  or  after  changing  the  “

Cartridge

”  or  every  6-12  months  depending  on  envi-

ronmental conditions. After entering the "

Zero Code

" as confirmation of the operation, there will be a flashing red Led 

and output becomes 4.0 mA. 

Note

: This function is not working if mA output is higher than the value shown in the table below, depending on the 

model, or if the red Led is already ON. In this case, please recalibrate the unit and/or replacing the “

Cartridge

” 

TS293EA-TS293EA-H 

TS293EC-S-TS293EC-

TS293ES 

TS293EH 

TS293EN 

TS293EN 

TS293EN2 

4.8 mA 

(15 ppm NH3) 

4.8 mA 

(15 ppm CO) 

6.4 mA 

(3 ppm SO

2

)

 

4.8 mA 

(5 ppm H

2

S)

 

8 mA 

(2.5 ppm HCN)

 

5.6 mA 

(10 ppm NO)

 

6.6 mA 

(5 ppm NO

2

)

 

“CALIBRATION” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the sensor 

using the gas mixture indicated above.

 

Warning

to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-

low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and 
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation. 

Important note:

 

during Calibration routine the mA output indicates 0mA

.

 

The “

Calibration

” can be done in clean air only (

environment without the presence of toxic or other polluting gas

). With the keys 

perform the “

Calibration Code

". Wait until the yellow and green Led switch on fix and the red Led starts to flash. In-

sert the TC011 on the sensor holder and adjust the influx of the gas the way that the flow meter reaches the value 
indicates in table below in 

column 3

, depending on the model (

see also Fig.3

). Wait around 3-5 minutes, until the red 

Led lights (

and while it is ON

), press the key F2 and hold it pressed until the red LED is switched off for at least 2 sec-

onds (

but if the red LED continues to blink more than 5-6 minutes means that the gas flow is not constant or unsat-

isfactory, or the concentration is not as required, stop the operation, turn off the instrument and recalibrate

). 

Then, 

close the gas cylinder and remove TC011

. At this point we can have two possibilities: 

Yellow  and  green  Led  illuminates:

  the  calibration  routine  is  correct. 

Wait  8  seconds,  until  the  instrument  auto-

matically restores the normal working conditions

(

See “

Operational Description > Preheating

). 

Yellow Led illuminates:

 the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-

peat 

Preheating

,  then  repeat  the  “

Calibration

  routine  without  inserting  again  the  code.  If  condition  still  persists  after 

the replacement of the “

Cartridge

”, please send the detector back to the manufacturer for reparation. 

“CALIBRATION CHECKS” (

no Code required

)

:

 

using the gas mixture indicated above, this function is used to con-

trol, the correct response to the gas and can be made after the "

Calibration

" or the installation, but should be done 

during the periodic maintenances, as it is the only method to verify the effective functioning of the detector.

 

Insert the TC011 on the sensor holder and adjust the sample gas bottle’s valve, so that the flow meter reaches the 

value  indicates  in  table  below  in 

column  3

,  depending  on  the  model  (

see  also  Fig.3

).  Check  that  the  voltmeter,  with 

probes connected on  tests points “

TEST mV

”, 

(

see fig.2

reaches (in  2÷3 minutes) the  value shown  in 

column 4

, de-

pending on the model, [

corresponding to 

mA

 output indicated in 

column 5

 and the Central Unit which is connected to the detector, dis-

plays  the  value  indicated  in 

column  6

].  If  the  result  is  different,  is  necessary  to  recalibrate  the  sensor  (

see  “

Calibration

). 

Then, close the gas bottle, remove TC011. The mA output will slowly decrease up to 4 mA (

normal working conditions

). 

 
 
 
 

Summary of Contents for TS293E Series

Page 1: ...de ZS EC2 EX TS293ES SO2 Biossido di zolfo Sulfur dioxide ZS ES1 EX TS293EH H2S Idrogeno Solforato Hydrogen sulphide ZS EH1 EX TS293EHCN HCN Acido cianidrico Hydrogen cyanide ZS EHCN EX TS293EN NO Ossido di azoto Nitric oxide ZS EN1 EX TS293EN2 NO2 Biossido di azoto Dioxyde d azote ZS EN2 EX Con Cartuccia Sensore Sostituibile Inside Replaceable Cartridge Sensor Avec Cartouche Capteur échangeable S...

Page 2: ...nni years ans 2 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi month mois 1 mese month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois Variazione dello Zero Zero shift Dérive du Zero 6 ppm 4 ppm 10...

Page 3: ... ha densità relativa all aria di 2 25 Il suo TWA è 2 ppm Time Weighted Average e lo STEL è 5 ppm Short Term Exposure Limit L SO2 è un forte irritante per gli occhi e le vie respiratorie Per inalazione può causare edema polmonare e una prolungata esposizione può portare alla morte Nel caso si combini con l acqua forma Idrogeno Solforato H2S TS293EH ha un sensore in grado di rilevare Idrogeno Solfor...

Page 4: ...ilitato il Led rosso e quindi tutti i Led sono accesi uscita 20mA indica o il guasto della Cartuc cia Sensore oppure una concentrazione di gas superiore al Fondo Scala Se non è presente alcuna fuga di gas e la condizione persiste anche dopo la sostituzione della Cartuccia sarà necessario inviare il rilevatore al fornitore per la riparazione INSTALLAZIONE I rilevatori vanno installati posizionati e...

Page 5: ... codice Kit di Taratura e Bombole con Miscela Gas titolato per Calibrazione e Verifica la miscela da utilizzare in funzio ne del modello installato è MODELLO Bombola di Gas Titolato Codice Bombola TS293EA TS293EA H 100 ppm Ammoniaca NH3 in azoto Fornibile a richiesta TS293EC S TS293EC H 300 ppm Monossido di carbonio CO in azoto BO210 TS293ES 10 ppm Biossido di zolfo SO2 in azoto Fornibile a richie...

Page 6: ...to in colonna 3 della tabella ovvero che l uscita in mA aumenti fino al valore indicato in colonna 4 e la centrale cui è collegato il rilevatore indichi circa il valore indicato in colonna 5 Se il valore fosse diverso effettuare la Calibrazione Terminata la Verifica chiudere la bombola e togliere il TC011 Attendere 3 5 minuti per tornare alle condizioni di funzionamento normale 1 Modello 2 Gas da ...

Page 7: ...ave effects on the blood causing the formation of metaemoglobina In more severe cases it can lead to death Nitric ox ide is a pollutant that is generated in part in combustion processes in part by emissions from natural volcanic erup tions fires or civil activities and industrial processes involving combustion such as transport vehicles with diesel en gine gasoline LPG etc and the production of he...

Page 8: ...NITS In case of a central unit other then Tecnocontrol please verify the max load resistor as shown in Fig 4 The max load resistor is 50 ohm with 12Vdc power supply while is 400 ohm with 24Vdc power supply We suggest using 24Vdc power supply WARNING Average life In clean air and in normal working condition the sensor s life is about 2 years After this period the yellow Led FAULT flashes every 4 se...

Page 9: ...2 F1 this function allows completely recalibrate the sensor using the gas mixture indicated above Warning to guarantee that no errors of elaboration happen the possibility exists that during the Calibration the yel low Led switch off every 8 seconds in this case interrupt the procedure switch off and switch on the instrument and repeat the Calibration If condition persists it will be necessary to ...

Page 10: ...dans le l industrie du froid pour la production d explo sifs pour la production de plastiques et polymères et pour les solvants Le TS293EA H est comme le modèle précédente mais il utilise un capteur adapté pour les environnements pol lués et de l industrie et à la présence d autres gaz toxiques Le TS293EC S permet de détecter le Monoxyde de Carbone CO qui est un gaz toxique inodore sans couleur es...

Page 11: ...onnement Normal seule la led vert ON doit être allumée Alarme Le led rouge ALARM s allume si la concentration de gaz dépasse la valeur indiquée dans le tableau ci dessous selon le modèle Abilitation avec Dip Switch n 1 ON TS293EA TS293EA H TS293EC S TS293EC H TS293ES TS293EH TS293EHCN TS293EN TS293EN2 50 ppm NH3 200 ppm CO 10 ppm SO2 20 ppm H2S 5 ppm HCN 25 ppm NO 30 ppm NO2 Dérangement La central...

Page 12: ...93EHCN TS293EN TS293EN2 500 ppm NH3 1000 ppm CO 150 ppm SO2 500 ppm H2S 100 ppm EHN 1000 ppm NO 150 ppm NO2 Attention Les capteurs électrochimiques contiennent une petite quantité d acide sulfurique qui est corrosif Nor malement il ne doit y avoir aucune perte au capteur dans le cas contraire ceci est du à une action mécanique vio lente ou à un usage hors limites de fonctionnement Dans le cas d un...

Page 13: ... dique la valeur indique dans le tableau ci dessous dans la colonne 3 voir aussi Fig 3 Attendre 3 5 minutes envi ron puis quand la led rouge s allume et pendant qu elle est allumée appuyer sur la touche F2 et maintenir cette pression tant que la led rouge ne reste pas éteinte durant au moins 2 secondes si en revanche la led rouge conti nue à clignoter plus de 5 6 minutes cela signifie que le flux ...

Page 14: ...ido di azoto Nitrogen dioxide dioxyde d azote 20 ppm 2 2 1 Cloro Clorine Chlore 1 ppm 0 0 0 Etanolo Ethanol Éthanol 200 ppm 0 0 2 5 0 Etilene Ethylene Éthylène 100 ppm 0 0 80 50 Idrogeno Hydrogen Hydrogène 100 ppm 0 0 2 46 40 Idrogeno Solforato Hydrogen sulphide Sulfure d hydrogène 10 ppm 2 0 2 0 0 5 0 3 4 Monossido di carbonio Carbon Monoxide Monoxyde de carbone 100 ppm 0 1 100 100 1 Ossido di az...

Page 15: ...l Unit Centrale de détection RL S Dip Switch Test Uscita Output Test Sortie Test 40 200mV TP1 TP2 mV S Uscita 4 20 mA 4 20 mA Output Sortie en 4 20 mA Fig 2 Schema di collegamento Wiring diagram Schéma Fig 3 Tester di calibrazione Calibration Tester Kit de Calibration 0 3 L min 1 L min Alimentazione Power Supply Alimentation 100 300 400 500 50 0 5 200 Vdc RL Resistenza di carico Load resistor Resi...

Reviews: