background image

IST-2138.EO01.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.6/11

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

 Tel: 02/26 92 28 90   Fax: 02/21 33 734

 

 

 

Note:

 the detector is not able to detect Oxygen gas outside the room where it is installed, neither inside walls nor 

under the floor. 

Sensor operating limits: 

the electro-chemical cell, in any case, should not be exposed to temperatures lower than 

-40° or higher than 60°C. The electrochemical cell should not be exposed to an over-range concentration of 30 % 
O

2

. If this condition occurs, please replace the “

Cartridge Sensor

”. 

Safety:

 Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this mate-

rial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of 
its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, carbon

 

and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be washed 
with a copious supply of water. 

Cross  sensitivity  to  other  gases:

  The  electrochemical  cell  employed  has  a  good  resistance  towards  products 

such  as  sprays,  detergents,  ammonia,  glues  and  paints. 

However,  in  case  of  products  containing  substances  in 

great quantity, these could interfere with the Sensor and cause false alarms.

 We recommend ventilating the room 

when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell 
needs one or more hours in clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY  ATTENTION

:  This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained  people; 

because starting this procedure it will start both Outputs (relays) causing the activation of connected alarm devices. 

Operation Check, Calibration in Air and Fine mA ADJ Output:

 are different code protected functions. To access 

these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pressure 

recognized, hold pressing it for around a second (

until the green Led doesn't switch off for a moment

). Then the next key can 

be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check

) please, only using a mixture

 

16% volume Oxygen (O

2

) in nitrogen 

You can use either the disposable cylinder, with adjust valve or

 

the high pressure ones with reduction gear. Is also 

necessary to use 

the calibration kit 

Tecnocontrol mod. TC011

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off and all 

the relays are deactivated. Then they will switch on in sequence, the Led, from the yellow up to the 3

rd

 red. To the 

lighting  of  the  various  Led  it  corresponds  the  activation  of  the  relevant  relays  (

relay  "FAULT"  with  the  yellow  Led,  relay 

"ALARM1" with the 1° red Led, relay "ALARM2" with the 2° red Led, relay "ALARM3" with the 3° red Led

). At the end all the Led will 

remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to 
perform this operation every 12 months according to the use. 

Note

: this function is not working if the 1° and/or the 2° and/or the 3° red Led have already turned on. 

“CALIBRATION IN AIR” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the 

sensor at 20.9%volume in air. This procedure when it is launched will automatically until the end. 

Warning

to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yellow 

Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and repeat 
the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation. 

Important note:

 

During Calibration routine the mA output indicates 0mA

.

 

The “Calibration in Air” can be done in clean air only (

environment without the presence of toxic or other polluting gas

). With 

the keys perform the “

Calibration

 Code

 ". Wait until the yellow and green Led switch on fix and the 1

st

 red Led starts to 

flash (the 3

th

 Led red

 

switch on to indicate the calibration working). Wait while the 1

st

 red Led switched off. At this 

point we can have two possibilities: 

Yellow and green Led illuminates:

 the calibration routine is correct. 

Wait 8 seconds, until the instrument automat-

ically restores the normal working conditions

(

See “

Operational Description > Preheating

). 

Yellow Led illuminates:

 the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-

peat 

Preheating

,  then  repeat  the  “

Calibration

  routine  without  inserting  again  the  code.  If  condition  still  persists  after 

the replacement of the “

Cartridge

”, please send the detector back to the manufacturer for reparation. 

Important

If the environment is much polluted and there is doubt that the concentration of oxygen is 20.9%v/v, you 

can use an air cylinder with 

20.9%v/v of O

2

 or use a gas balloon (

our code 

PA020

 ) filled with fresh air. In this case, 

before executing the "

Calibrating Code

", insert the TC011 on the sensor holder, adjust the flow of air, so that the flow 

meter indicates approximately 0.3 l/min (

see Fig.3

). 

"

Fine mA ADJ Output”

 (

ADJ Code: F2, F1, F1, F2

): This feature allows you to adjust the mA output of ± 0.32 mA cor-

responding to ± 0.5% O

2

, but it should be used only at new installation and if, after the "

Calibration in Air

" (clean air), 

the control unit, connected to SE138EO, indicates a value other than 20.9% O

2

After entering the "

ADJ 

Code

", the yellow Led starts to flash, the output value increases pressing the F1 key, and 

decreases pressing F2. Hold down the two F1 and F2 keys simultaneously for 2 seconds to confirm and then check 
that yellow Led remains off. This procedure cannot be used, if the yellow Led is already on, or if the O

2

 concentra-

tion is less than 20.6% or higher to 21.2%. If this is the case: "

Calibration in Air

" routine must be done first. 

“CALIBRATION  CHECK”  (

no  Code  required

)

:

 

although  the  sensor  operation  is  guaranteed  if  it  is  regularly  per-

formed successfully "

Calibrating in Air

", it is possible to simulate the response of the detector using sample gas cylin-

Summary of Contents for SE138EO

Page 1: ... linéaire Campo di misura Standard Range Champ de mesure 0 25 0 O2 Limite del Sensore Sensor limits Limite échelle 30 O2 Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi 3 month 3 mois Tempo di risposta Response Time Temps de réponse T90 30 secondi seconds secondes Ripetib...

Page 2: ... di stabilizzazione del sensore segnalata dal lampeggio del Led giallo FAULT Dopo circa 60 secondi il Led giallo si spegne e si accende il Led verde ON che indica il normale funzionamento Dopo questo tempo il sensore è in grado di rilevare il gas ma raggiunge le condizioni di stabilità ottimali dopo circa 15 minuti di funzionamento continuo Funzionamento Normale deve essere acceso il solo Led verd...

Page 3: ...eggiarsi anche se è esposto a concentrazioni di O2 molto elevate oltre il 30 In questi casi il sensore deve essere sostituito Attenzione i sensori elettrochimici per Ossigeno contengono una piccola quantità di acetato di potassio che è cor rosivo Normalmente non si verificano perdite dal sensore questo può essere causato solo da un azione meccanica violenta o un uso oltre i limiti di funzionamento...

Page 4: ...uccesso la Calibrazione in Aria è comunque possibile controllare la corretta risposta del rilevatore al gas uti lizzando la miscela di gas sopra indicata Infilare il TC011 sul portasensore regolare il riduttore della Bombola in modo che il flussometro indichi circa 0 3 l min vedi Fig 3 Attendere circa 1 2 minuti e controllare che il rilevatore vada in allarme come decritto nel capitolo Funzionamen...

Page 5: ...device If the condition does not change it will be necessary to replace the unit and or send it back to the supplier to repair Yellow and green Led activates FAULT relays activate and 0mA output signal this happens when the Cartridge Sensor is not working First try to perform the procedure of Calibration in Air as described in the section Test and Calibration Calibration in Air finally try to repl...

Page 6: ...Led relay ALARM2 with the 2 red Led relay ALARM3 with the 3 red Led At the end all the Led will remain lighted for around 5 seconds then the central returns at the conditions of normal operation It is advisable to perform this operation every 12 months according to the use Note this function is not working if the 1 and or the 2 and or the 3 red Led have already turned on CALIBRATION IN AIR Calibra...

Page 7: ...ment normal Led jaune FAULT capteur en panne ou déconnecté ou saturé ou échu Touches F1 et F2 pour les opérations de vérification et calibrage CARACTÉRISTIQUES DU GAZ Le SE138EO permet de détecter l Oxygène O2 qui est un gaz incolore inodore légèrement plus lourd que l air sa densité relative à l air est de 1 1 Ce n est pas un gaz inflammable mais étant fortement oxydant il réagit avec les matéria...

Page 8: ... plancher l oxygène ayant une densité proche de celle de l air Raccordements électriques Fig 2 la distance maximale à laquelle chaque détecteur peut être alimenté est indi quée dans le Tableau 1 en fonction de la section du câble utilisé Dans le cas d alimentation de plusieurs détecteurs en parallèle il est nécessaire de calculer la chute de tension sur les portions communes des câbles Un câble sa...

Page 9: ... à à la fin Durant la procédure une rare possibilité existe que la led jaune clignote chaque 8 secondes dans ce cas inter rompre la procédure puis exécutez le contrôle en coupant et en rétablissant l alimentation du détecteur En cas de persistance il sera nécessaire de remplacer et ou renvoyer le détecteur au fournisseur pour réparation Important Durant l opération de étalonnage l appareil portant...

Page 10: ...l alimentatore a 12Vcc The maximum distance to install each detector from the 12Vdc power Supply Distance max d installation d un détecteur sous 12Vcc La max distanza cui può essere installato ogni rivelatore dall alimentatore a 24Vcc The maximum distance to install each detector from the 24Vdc power Supply Distance maxi d installation d un détecteur sous 24Vcc 0 75 mm2 26 Ω km 100 m 300 m 1 mm2 2...

Page 11: ... Output Board Sortie sur Realis 12 24Vdc Alimentazione Power Supply Alimentation Centrale rivelazione gas Gas Central Unit Centrale de détection RL Uscita in 4 20 mA 4 20 mA Output Sortie en 4 20 mA Test Uscita Output Test Sortie Test 40 200mV mV 0 3 L min Fig 3 Tester di calibrazione Calibration Tester Kit de Calibration Alimentazione Power Supply Alimentation 100 300 400 500 50 0 5 200 Vdc RL Re...

Reviews: