background image

Overall accuracy: +/- 0,5 % fs
Sampling rate: 1 sample per second
Action: 1C type according to EN 60730-1
Compliance: ECC directive EMC 89/336 (EN 50081-1, EN 50082-1), ECC
directive LV 73/23 and 93/68 (EN 60730-1)

MEASUREMENT RANGE TABLE

-99.9 ... 700.0

- 999 ... 7000

4..20 mA, 0..20 mA,

0..10 V

 

(gener.)

---

0 ... +1600 °C

+32 ... +2912 °F

tc S (rh)

---

0 ... +1200 °C

+32 ... +2192 °F

tc K (Cr)

---

0 ... +800 °C

+32 ... 1472 °F

tc J (FE)

-50.0 ... +150.0 °C

-58.0 ... +302.0 °F

-50 ... +150 °C

-58 ... +302 °F

Ni 100 (ni)

-99.9 ... +600.0 °C

-99.9 ... +999.9 °F

-100 ... +600 °C

-148 ... +1112 °F

Pt 100 (Pt)

-50.0 ... +150.0 °C

-58.0 ... +302.0 °F

-50 ... +150 °C

-58 ... +302 °F

PTC

4 DIGIT with D.P.

4 DIGIT 

PROBE

3 - INSTALLATION
MECHANICAL MOUNTING

The instrument, in  DIN case 24 x 48 mm,

is designed for panel mounting. Make an hole 22 x 45 mm and insert the
instrument, fixing  it with the provided special brackets .We recommend to
mount the  gasket to obtain an IP 54 front protection. Avoid to place the
instrument in areas with  humidity or dirt. Connect the instrument as far as
possible from source of electromagnetic disturbances so as motors, power
relays, relays, electrovalves,etc. 

ELECTRICAL  CONNECTIONS

: Carry out the electrical wiring

connecting only one wire for each terminal , according to the following
diagram, check that the power supply is the same as indicated on the
instrument and the loads current is not upper than the maximum current
admitted. The instrument, being a built in equipment with permanent
connection into a cabinet, is not furnished with internal device protecting  
from  overcurrent : it's recommended , therefore, to properly protect all the
electric circuits connected to the instrument, with devices (ex. fuses)
proportionate to the circulating currents. It's strongly recommended to use
cables with proper insulation, according to the working voltages and  
temperatures. Furthermore, the input cable of the probe has to be kept
separate from line voltage wiring. If the input cable of the probe is
screened, it has to be connected on the ground with only one side. It is
advisable to check that the parameters are those desired before connecting
the outputs to the actuators so as to avoid malfunctioning . Whenever a
failure of the instrument could cause dangerous or damaging situations, it
should be kept in mind that the plant has to be provided with additional
devices to ensure the safety.

4 - OPERATING MODE

4.1 - ON/OFF CONTROL

The control mode of the instrument by ON/OFF mode occours on  the
outputs according to the Sets point fixed, to the differential switching
points (par. "d1", "d2") and to operating modes (par. "HC1", "HC2")
programmed. For a correct functioning it is necessary to set a negative
differential switching point for reverse controls (ex.Heating processes) and
a positive differential switching point for direct controls (ex.Cooling
processes). Furtheremore the two Sets can be programmed independently
or dependently throught the parameter Sets Connection ("OCO"), and in
the case of set dependentlythe real regulation Set of output 2 will be
[Set1+Set2]. If output 2 is used as an alarm output, par. "OCO"
determines if the alarm is absolute or relative (in=absolute, di=relative)
and par. "HC2" determines if the alarm is low or high (H=low, C=high).
The operating mode can be also modified by means of outputs delay
parameter (par. "od") or  Dynamic Set Point  function, see the next
chapters for these functions.

4.2 - NEUTRAL ZONE CONTROL

The Neutral Zone control mode concerns both outputs and is generally
necessary for controllig a plant that possesses an Positive regulation
element (ex. Heating, Humidification etc.) and a Negative regulation
element (ex. Cooling, Dehumidification etc.) . The functioning is
determined by the Set1 (Set2 is automatically disconnected) and Neutral
Zone (par. "db"). Subsequently the regulator operates as follows: switches
off the outputs when the process value is the same as Set, switch on the
output OUT1 when the process value is under as [Set1-db], and switch on
the output OUT2 when the process value is upper as [Set1+db]. Thus the
Positive regulation element will be connected to output OUT1, while the
Negative regulation  element is at output OUT2.

4.3 - PID CONTROL AND AUTOTUNING FUNCTION

PID control mode can only be carried out through output OUT 1 while
output OUT 2 can only operates by means of ON/OFF control mode with
Set 2 independent or dependended in respect with Set 1 and can therefore
be used as an alarm (see ON/OFF regulator). The PID algorithm of the
instrument is provided to program the following parameters:
For PROPORTIONAL term :

"Pb"

 - Proportional Band,

"rst"

 - Manual reset

"Ct"

 - Cycle time.

For INTEGRAL term:

"It"

 - Integral time

For DERIVATIVE term:

"dt"

 - Derivative time

The instrument is set on parameters relative to PID control of standard
value. If ever these should result to be unsuitable it would be advisable to

TECNOLOGIC - THP 24

 

 USER MANUAL (I - GB) - Vr. 01   - ISTR 00239  - PAG. 6

Summary of Contents for THP 24

Page 1: ...l The herewith issue is an exclusive property of TECNOLOGIC S p A which forbids any reproduction and divulgation although partial if not expressly authorised TECNOLOGIC S p A reserves the right to execute aesthetically and functional modifications at any moment and without any notice INDICE PROBLEMI MANUTENZIONE E GARANZIA 7 TABELLA PARAMETRI 6 1 DESCRIZIONE DEI PARAMETRI 6 IMPOSTAZIONE DEI LIMITI...

Page 2: ...DIGIT with D P 4 DIGIT PROBE 99 9 700 0 999 7000 4 20 mA 0 20 mA 0 10 V gener 0 1600 C 32 2912 F tc S rh 0 1200 C 32 2192 F tc K Cr 0 800 C 32 1472 F tc J FE 50 0 150 0 C 58 0 302 0 F 50 150 C 58 302 F Ni 100 ni 99 9 600 0 C 99 9 999 9 F 100 600 C 148 1112 F Pt 100 Pt 3 INSTALLAZIONE MONTAGGIO MECCANICO Lo strumento in contenitore DIN 24 x 48 mm è concepito per il montaggio ad incasso a pannello P...

Page 3: ...mento visualizzerà EEE Per interrompere il ciclo di Autotuning o ripristinare il funzionamento normale dopo un errore spegnere e riaccendere lo strumento I valori calcolati saranno memorizzati automaticamente dallo strumento al termine dell esecuzione del ciclo di Autotuning nei parametri relativi alla regolazione PID 4 4 FUNZIONE SET POINT DINAMICO La funzione Set Point Dinamico agisce solo sull ...

Page 4: ...Offset positivo o negativo che viene sommato al valore letto dalla sonda prima della visualizzazione alla quale è legato anche il funzionamento di regolazione Questo parametro può rendersi necessario per una ritaratura dello strumento Ft TIPO DI REGOLAZIONE Permette di selezionare uno dei possibili modi di regolazione che lo strumento offre ON OFF on PID Pi o Zona Neutra nr PSE SONDA IN INGRESSO P...

Page 5: ...y kind of probe having normalised output in current 4 20 mA 0 20 mA or voltage 0 10 V 1 1 FRONT PANEL 1 Key P Used for the set point setting and to program the functioning parameters 2 Key DOWN Used to decrease the values or to select parameters 3 Key UP Used to increase the values or to select parameters 4 Led OUT1 Signalize when the output OUT1 is on on or off off and signalize the set point 1 o...

Page 6: ... OFF CONTROL The control mode of the instrument by ON OFF mode occours on the outputs according to the Sets point fixed to the differential switching points par d1 d2 and to operating modes par HC1 HC2 programmed For a correct functioning it is necessary to set a negative differential switching point for reverse controls ex Heating processes and a positive differential switching point for direct c...

Page 7: ...lue Once the desired value has been set release P and the selected parameter code can be read on the display By pressing UP or DOWN it is therefore possible to choose another one and modify it as previously mentioned To outgoing from the programming mode no key is to be pressed for about 20 seconds the instrument will automatically return to normal functioning mode visualizing the process value Th...

Page 8: ...5 and 9 For example if the real measure is 78 display will show 75 or if the real measure is 70 3 the display will show 70 0 n without approximation y with approximation tAb FIXED PARAMETER 6 1 PARAMETERS TABLE 0 0 Probe limit Restart band Dynamic Set drb 20 1 500 sec Cycle time Ct 0 0 500 sec Outputs delay od 0 Probe limit Manual reset rSt 30 0 3600 sec Derivative time dt 300 0 3600 sec Integral ...

Reviews: