background image

Fig. 1. Simplified schematic of OPT-1x20 internal circuits.

As you can find out from the schematic... 
• the analog input can be equipped with a current shunt, or the input impedance

can be reduced by replacing R1 (a resistor with value of 1 MOhm and tolerance
of 1% is fitted at the factory);

• the input amplifier is equipped with positions for two dividers determining the

gain; the first is reserved for users and the factory setting is 1x amplification 
(i.e. a resistor with value of 0 Ohm is fitted in position R5 and position R6 is left
free), the second is factory fitted with a divider for 10x gain (18K and 2K
resistors with a tolerance of 0.1%);

• after changing the range (i.e. relocating the JP1 jumper position) it is

recommended to re-compensate the offset and calibrate the gain (see the meaning
of potentiometers P1 and P2).

Warning: The manufacturer is not liable for damage to the OPT-1x20 board or

other damage caused by additional installation or change of components.

to the DAQ card

input

gain
adjustment

1x

offset

P1

P2

1M0

- - -

R5 0R

2K0

18K

R7

R6

R8

R1

isolation

10x

1x

adjustment

range configuration

JP1

Manufacturing, sales office, service center and technical support: 

TEDIA spol. s r. o.
Zabelska 12, 31200 Plzen, Czech Republic

phone/e-mail:

https://www.tedia.eu/contacts

address:

OPT-1020/1120

User Guide

(further information available at http://www.tedia.eu)

document revision: 02.2013, © 1994-2013 TEDIA

®

 spol. s r. o.