background image

4. Turn on the instrument and use the following table to verify that 

the readings on each range are within specifications.

Table 1. Performance Verification

Model No.  Range     Reading     Model No.  Range   Reading

874F         200      32.0 ± 1.2°F       879         760      0 ± 1°C

2000        32 ± 2°F                       1400    32 ± 2°F

874C         200      0.0 ± 0.5°C        875F       200     32 ± 1.2°F

1100        0 ± 1°C                        752     32 ± 2°F

878          1100        0 ± 1°C          875C        200   0.0 ± 0.6°C

2000       32 ± 2°F                        400      0 ± 1°C

BATTERY INSTALLATION/REPLACEMENT

WARNING

Turn the instrument off and disconnect the

input probe before removing the bottom

cover. Reinstall the cover before resuming

use of the instrument.

1. Place the unit face down on a bench or other similar surface 

and remove the screws from the bottom cover.

2. Separate the bottom cover from the rest of the instrument by

grasping the top of the case (just above the display) and care-

fully lifting it away from the display.

3. Remove the old battery.

4. Place the new battery in the battery compartment. Be sure to 

observe the proper polarity. Refer to Figure 2.

5. Reinstall the bottom cover before resuming use of the instru-

ment.

Figure 2 Battery Installation

6

NOTE

An open sensor or overrange condition is indicated by the

following display.

3

Figure 1.

Model 878 Control Layout

All Others Similar

On/Off Range Switch

3 1/2 Digit Display

Minus Sign Displayed

Plus Sign Implied

Low Battery Indicator

Miniature TC Connector

i

First digit displays
the number one.

Remaining display
blank.

BATTERY

BATTERY TERMINALS

Summary of Contents for Budget 874F

Page 1: ...anufactured by Elcto Scientific Industries ESI Gertsch Keithley Instruments Lucas Weinschel and Pragmatic Instruments A complete list can be viewed on our Product Service Directory at www tegam com For more information about TEGAM and our products please visit our website at www tegam com or contact one of our customer service representatives at sales tegam com or 800 666 1010 Ten Tegam Way Geneva...

Page 2: ... to be defective To exercise this warranty write or call Tegam Inc in Geneva Ohio You will be given prompt assistance and return instruc tions Send the instrument transportation prepaid to the indi cated servive facility Repairs will be made and the instrument returned transportation prepaid Repaired products are war ranted for the balance of the original warranty period or at least 90 days LIMITA...

Page 3: ...n materials Has a one foot 5 extended coil cord with a molded plastic handle Accuracy is 2 2 C or 0 75 whichever is greater Time con stant is 1sec stir red liquids Max tip temperature is 200 C or 392 F Model 8756 Compact Air Gas Probe for measurement of air gasses Exposed junction within protective shroud responds to ambient temperature while shielded from incident radiation Has a one foot 5 exten...

Page 4: ...1A Thermocouple Kit includes the Model 8712 Thermocouple Sensor 6m 20 of AWG 24 Type K duplex TC wire a mininature TC plug and jack and a standard TC plug and jack Model 8712 Thermocouple Sensor is a three foot 24 gauge TC wire sensor that has a beaded thermocouple junction Accuracy is 2 2 C or 0 75 whichever is greater Time constant is 1sec Max tip temperature is 480 C or 900 F Model 8713 Regular...

Page 5: ...er similar surface and remove the screws from the bottom cover 2 Separate the bottom cover from the rest of the instrument by grasping the top of the case just above the display and care fully lifting it away from the display 3 Remove the old battery 4 Place the new battery in the battery compartment Be sure to observe the proper polarity Refer to Figure 2 5 Reinstall the bottom cover before resum...

Page 6: ...tion contains information needed to maintain your instrument The following information is included functional check performance veri fication and battery installation replacement WARNING The information presented in this section is intended for use by qualified personnel only Turn off the power and remove all probes from the instrument before replac ing the battery FUNCTIONALITY CHECK At room temp...