DMM4040/4050
Technical Reference Manual
AC Current Verification Steps
Connect the Meter to the test equipment as shown in Figure 3-9 and, depending on which
meter you are calibrating, apply the nominal values listed in Table 3-14. Verification
forms can be found in Appendix A which can be copied and used to record each meter
reading.
Note
The 400 mA range is only available in Outguard version 2.0 and later.
gdc022.eps
Figure 3-9. AC Current Equipment Setup
3-24