Load the Filter Software

1.

Insert the CD in the oscilloscope drive.

2.

Navigate to the CD folder and make a copy
of the Socket Cable XL Filter Files.

3.

Paste the

fi

les in C:/TekScope/

Math_Arbitrary_Filters/Probe-Filters.

4.

From the Vertical menu, click Probe Cal.
Under Probe Tip, select Other Tip.

5.

From the Math menu, select Math Setup.

6.

Click Editor and select Filter.

7.

In the user-de

fi

ned Arbitrary Filter area,

click the Load button.

8.

Navigate to the directory that you stored the

fi

lter in (from step 2) and select the

fi

lter.

Using the Filter Software

Two

fi

lters are available:

XL_HTT_DIFF_Multirate.

fl

t

(for differential mode)

XL_HTT_A_B_COMM_Multirate.

fl

t

(for A-Gnd, B-Gnd, and CM modes)

Both

fi

lters are multi-rate

fi

lters that can

operate at sampling rates from 50 GS/s down
to 1.25 GS/s. Since these are multi-rate

fi

lters,

the

fi

lter bandwidth and rise time vary with the

sample rate.
The

fi

lters are 4000 points long, so your

oscilloscope record length must be at least 2X
or greater to display a usable waveform. The

fi

rst and last 2000 points of record length are

truncated from the

fi

ltered waveform.

More information about using

fi

lters is located in

the Tektronix Digital Oscilloscopes Online Help
in the “Math Equation Editor (Filter Tab)” topic.
For speci

fi

c information on arbitrary

fi

lters, refer

to “Arbitrary

fi

lters, math”.

Replacing the Resistors

The resistors that are presoldered to the solder
tips need to be replaced periodically, due to
normal wear. A kit of replacement resistors is
available (order part number 020-2937-00).

1.

Apply a soldering iron to the via on the tip
board and remove the existing wire. Work
quickly with a low-heat soldering iron.

If a small piece of wire is in the via, heat the
solder and use another piece of wire to push
the wire fragment from the via.

2.

Add fresh solder to the via, if necessary.

3.

Cut one lead of the resistor to 0.2" (5mm).

4.

Heat the solder in the via and thread the cut
resistor lead into the via, leaving enough
lead length to bend the resistor

fl

at with the

tip board.

5.

Cut excess lead on the bottom of the board.

6.

Cut the remaining resistor lead as necessary
to reach your test point.

Dimensions

Electrical Characteristics (Typical)

Table 1: Socket Cable

1

Tip/measurement
mode

Rise time, ps

10–90% 20–80%

BW,
GHz

TriMode High Temp Solder Tip

Differential (A-B)

60

40

6

Single-ended (A, B)

120

80

3

Common-mode

150

100

2.5

TriMode Micro-Coax Solder Tip

Differential (A-B)

100

60

4

Single-ended (A, B)

150

100

2.5

Common-mode

200

135

1.5

Damped Wire Pair

Differential (A-B)

50

35

8

1

Temperature range 0 °C to +40 °C (32 °F to 104 °F)

Table 2: Socket Cable XL with TriMode High Temp
Solder Tip and DSP

fi

lter

fi

les

1 2

Measurement mode

Differential

A, B, Com

Acquisition
rate, Gs/s

Rise time,
ps, 10–90%

BW,
GHz

Rise time,
ps, 10–90%

BW,
GHz

50

55

10

150

3

25

60

8

150

3

12.5

120

4

150

3

6.25

230

2

230

2

3.125

450

1

450

1

1.5625

900

0.5

900

0.5

1.25

1150

0.4

1150

0.4

1

The DSP

fi

lter

fi

les must be used with the Socket Cable XL.

2

Temperature range –55 °C to +150 °C (-67 °F to 302 °F)

Warranty Information

For warranty information, go to
www.tektronix.com\warranty.

Contacting Tektronix

Tektronix, Inc.
14200 SW Karl Braun Drive
PO Box 500
Beaverton, OR 97077
USA

For product information, sales, service, and
technical support:

In North America, call 1-800-833-9200.

Worldwide, visit www.tektronix.com to

fi

nd

contacts in your area.

Copyright © Tektronix, Inc. All rights reserved.

www.tektronix.com

Reviews: