Getting Started
4
P7350 5 GHz Differential Probe Instruction Manual
NOTE
. If your instrument supports probe calibration routines, now is
a good time to perform them.
6.
Disconnect the probe from the PROBE COMPENSATION
connector, and connect the two KlipChips together.
7.
With the probe offset set to 0.0 V, the oscilloscope display should
be at the ground reference.
8.
Set the oscilloscope volts/division to 500 mV.
9.
Adjust the probe offset. The displayed waveform should vary
between approxi1.25 V and --1.25 V.