P7350 5 GHz Differential Probe Instruction Manual
21
Appendix A: Specifications
The specifications in Tables 3 through 5 apply to a P7350 probe
installed on a TDS6604 oscilloscope. The probe must have a
warm-up period of at least 20 minutes and be in an environment that
does not exceed the limits described in Table 3. Specifications for the
P7350 differential probe fall into three categories: warranted, typical,
and nominal characteristics.
Warranted Characteristics
Warranted characteristics (Table 3) describe guaranteed performance
within tolerance limits or certain type-tested requirements.
Warranted characteristics that have checks in the
Performance
Verification
section are marked with the
n
symbol.
Table 3: Warranted electrical characteristics
Characteristic
Description
n
Rise time (probe only)
100 ps, +20
_
C to +30
_
C (+68
_
F to +86
_
F)
,
250 mV step
n
DC gain
0.16
±
2% (corresponds to 6.25 X attenuation)
n
Output offset voltage
±
10 mV +20
_
C to +30
_
C (+68
_
F to +86
_
F)
Maximum nondestructive input voltage
±
15 V(DC + peak AC) between signal and
common of the same channel.
Delay variation (probe-to-probe)
600 ps maximum
Temperature
Operating: 0 to +40
_
C (+32 to +104
_
F)
Nonoperating: --55 to +75
_
C (--131 to +167
_
F)
1