Appendix B: Performance Verification
P7350 5 GHz Differential Probe Instruction Manual
31
TekConnect Interface Calibration Adapter
To complete the performance verification and adjustment procedures
for your probe, you need to order the optional TekConnect Interface
Calibration Adapter, Tektronix part number 067-0422-00 (see
Figure 18). The adapter connects between the host instrument and
the probe under test and provides connectors for internal probe
measurements.
Figure 18: TekConnect Interface Calibration Adapter
When the adapter is connected to the oscilloscope, the adapter is
identified as a valid calibration device. However, additional power
supplies necessary to power the probe are not enabled until a
TekConnect probe is connected to the adapter and identified by the
oscilloscope. When a probe is detected through the adapter, the
Volts/div readout on the oscilloscope displays ##.
Refer to Table 8 on page 32 for detailed features of the calibration
adapter.