Appendix B: Performance Verification
34
P7350 5 GHz Differential Probe Instruction Manual
5.
Using the fine position and/or pressure adjust, maneuver the
probe so that the pins contact the CM or DM test points,
depending on which test you are performing. (See Figure 20.)
Common mode (CM)
Differential mode (DM)
Figure 20: Probe Calibration Fixture test points
6.
Verify that contact is made on both pins. (You may need to
readjust the fine position and/or pressure adjustment to make
positive contact with the test points.)
7.
Proceed with the specific test instructions.
Equipment Setup
Use this procedure to set up the equipment to test the probe.
1.
Connect the probe calibration adapter to the oscilloscope.
2.
Connect the probe to the probe calibration adapter.
3.
Turn on the oscilloscope and enable the channel.
4.
Allow 30 minutes for the equipment to warm up.