Appendix B: Performance Verification
P7350 5 GHz Differential Probe Instruction Manual
39
3.
Select the channel you have connected to on the 80E04 sampling
head, and then set the oscilloscope vertical scale to 50 mV/div.
NOTE
. The output of the step generator rises from a --250 mV level to
ground.
4.
Adjust the oscilloscope horizontal and vertical position controls
to display a signal similar to that shown in Figure 23.
5.
Set the oscilloscope horizontal scale to 50 ps/div and center the
waveform.
6.
Use the oscilloscope measurement capability to display rise time.
Increase the stability of the pulse edge measurement by using
averaging, if available. Rise time is determined from the 10% and
90% amplitude points on the waveform. Record the rise time
as
t
s.
The following steps instruct you to assemble the test setup that
includes the probe, as shown in Figure 24. The system and probe rise
time (
t
s+p
) that you measure in step 17 is used to calculate the probe
rise time (
t
p
) in step 18.
7.
Set the step generator control switch to standby.
8.
Remove the TekConnect-SMA adapter from the test setup.
9.
Connect the probe to the TekConnect calibration adapter.
10.
Connect the probe cal fixture to the step generator remote head,
and the termination to the other input of the probe cal fixture.
11.
Secure the probe head in the probe positioner.
The test setup should now be connected as shown in Figure 24 on
page 40.