Appendix B: Performance Verification
40
P7350 5 GHz Differential Probe Instruction Manual
SMA cable
Probe cal fixture
067-0419-00
CSA8000/TDS8000
TDS7404 Oscilloscope
Generator
remote head
Internal
clock output
Calibration
step generator
Trigger
input
80E04
Sampling head
extender cable
SMA
M-to-M
adapter
P7350 Probe
TekConnect
calibration
adapter
Figure 24: Test system rise time setup with probe
12.
Set the step generator control switch to on.
13.
On the TDS8000, expand the horizontal scale to help locate the
step edge in step 14, then adjust horizontal range to 500 ps/div
while maintaining the edge view. For a more stable measurement
display, turn averaging on.
14.
Using the probe positioner, probe the DM test points on the probe
calibration fixture. Compare your display to Figure 25 on page 41
to verify that you have a valid connection with both pins.