General
Information—7A42
SPECIFICATION
The
electrical
characteristics
listed
in Table
1-2
apply
when
the following
conditions
are
met: (1)
Adjustment
of
the
instrument
must
have
taken
place
at
an
ambient
temperature
between
+20°
and
+30°
C, (2)
the
instrument
is
allowed
a
20-
minute
warm-up
period,
(3)
specifications
are
valid
at
an
ambient temperature
of
0°
to +50°
C,
unless
otherwise
stated,
(4)
the
instrument
must
be
in
an
environment that
meets
the
limits
described
in
Table
1-3,
(5)
the
instrument
must
be
operated
in
a
calibrated
7000-series
mainframe.
Any
applicable conditions
not listed
above
may
be
stated
as
part of the
characteristic.
Environmental
characteristics
are
listed
in
Table
1-3
and
Physical
characteristics
are
listed in
Table
1-4.
TABLE 1-2
Electrical Characteristics
Characteristic
Performance
Supplemental
Requirement
Information
DISPLAY
Deflection Factor
Calibrated
Range
at
Input
BNC
Connector
TTL
(CMOS)
Family
0.1, 0.2,
0.5
V/div.
ECL
Family
20,
50, 100 mV/div.
Calibrated
Range
through
a
10X
Probe
TTL
(CMOS)
Family
1,
2,
5
V/div.
ECL
Family
0.2, 0.5,
1
V/div.
Channel
to Channel
Gain
Match
Within
2%
in
ECL
Logic
Family,
20
mV/
div,
1
Megohm
input
impedance.
1-6