Operating
Instructions—
7A42
NOTE
The
front-panel controls, indicators and connectors drawing,
Figure
2-7
in
this section of
the manual, can be unfolded and referred to
while
reading
the following text.
POSITION
A
POSITION control
is
provided
for
each of
the
four
channels. The
control
vertically
positions
the
displayed
trace
on
the
host mainframe crt
(clockwise
rotation
moves
the
trace upward).
GAIN
This
screwdriver
control
adjusts
the
7A42
display
output (of
all
four
channels)
to
match
the
vertical
gain
tolerance of
any
Tektronix 7000-series
mainframe.
STATUS INDICATORS (CH1, CH2, CH3, CH4)
Each
channel
has
four
status
indicators
(DISPLAY,
ECL/TTL,
GND
and
50Ω/1MΩ).
These
indicators
show
the
status of each
individual
channel.
A
description
of
each
indicator
is
given
below.
DISPLAY
When
the
DISPLAY
indicator
is
lit,
the
associated
channel
is
displayed
on
the
mainframe
crt.
If
the
indicator is
extinguished,
any
signal applied
to
the
channel
input
will
still
be
routed
to
the
TRIGGER
FUNCTION
circuitry,
providing
the
GND
indicator
is
not
lit.
ECL/TTL
When
the
ECL/TTL
indicator
is
lit,
the
channel
threshold
voltage
range
and
deflection
factors match
the
voltage
levels
and
signal
amplitudes of
the
TTL
logic
family.
When
the
indicator
is
extinguished
the threshold voltage
range
and
deflection factors
are
in
accord
with
the
ECL
logic
family.
GND
When
the
GND
indicator
is
lit,
the
input
to the
selected
channel
amplifier
is
grounded
and
the
signal
path
from
the
front-panel
bnc
connector
to
the
amplifier
is
open.
An
external
signal
applied
to
this
channel
is not
terminated
(it
is
open).
Grounded
channel
traces
are
still
displayed
to
enable
the
ground
reference
position
to
be
established.
50Ω/1MΩ
When
the
50Ω/1MΩ
indicator is
lit,
the
input impedance
of
the
associated
channel
is
one
megohm.
When
the
indicator
is
extinguished
the
input
impedance
is
50
ohms.
2-59