7B53A/7B53AN OPTION 5
b.
Check
for a test oscilloscope display of one or more
negative-going
pulses
in a group during the
vertical sync
pulse
interval. The
number of pulses is dependent on
the
television
standard
composite video
signal being applied
to
the
plug-in
unit. The pulse is very fast; approximately
0.1
µs
in width
at the 50%
amplitude level. Full pulse
amplitude
is approximately 2.8 volts.
4.
Check
TP920 Composite
Sync
Waveform
a.
Connect the 10X probe to TP920.
b.
Check test oscilloscope display. The
waveform should
be
negative-going composite sync occurring at the H rate;
amplitude
should be approximately 180 mV. All video,
except
burst, should
be clipped off.
5. Check
Sync
Separator Operation
a.
Apply
various television standard composite video
signals
(domestic and closed circuit) to the plug-in unit.
NOTE
If desired,
apply only those composite video signais
that
will be
used at your facility.
b. With
the 10X probe
connected to TP920, check the
test
oscilloscope
display for a
proper waveform as described
in step 4.
d.
Pull
the DLY TIME knob
outward and set the DLY
TIME
switch to the desired
position
(10µs). Rotate the
DELAY
TIME MULT control
to intensify the desired
portion
of
the display. Then, push the DLY TIME knob in
to view
the intensified
portion.
NOTE
In
this mode of operation the delayed sweep is free
running.
7.
Check
Discrete Line
Selector
Operation
a.
Set
the 7B53A/7B53AN
OPTION 5
Main Triggering
Amplifier
SOURCE switch to TV LINE. Adjust the LEVEL
control
to
obtain a triggered display.
b. Rotate the
DLY'D TRIG control out of the RUNS
AFTER
DLY TIME
position
and adjust
this control to
obtain
a triggered delayed sweep. Rotate the DELAY TIME
MULT
control to
display the desired portion of the signal.
NOTE
in
this mode of operation the
displayed portion of
the
signai will jump
from line to line as the DELAY
TIME
MUL Tcontrol
is slowly rotated.
c.
Disconnect the 10X
probe.
6.
Check
Variable Line
Selector Operation
a.
Apply
the
desired composite video
signal to the
vertical
plug-in
unit.
b. Check
that the 7B53A/7B53AN OPTION 5 Main
Triggering
Amplifier SOURCE switch
is
set to TV
FIELD
and
the
TIME/DIV switch
is set to 2 ms. Set the TIME/DIV
VARIABLE
control
to
display
slightly less than two fields.
Adjust the
Main Triggering LEVEL
control to obtain a
triggered display.
c. Set the
7B53A/7B53AN
OPTION 5 DLY'D
TRIG
control
to RUNS
AFTER DLY TIME position. Set the
Delayed Sweep Triggering controls as follows:
SLOPE
COUPLING
DC
SOURCE
TV
LINE
8. Check
Main
Triggering Amplifier TV Line Oper
ation
a.
Set
the TIME/DIV to 10 µs. Check
that the Main
Triggering
Amplifier SOURCE switch
is set to TV
LINE.
b.
Adjust the
Main Triggering Amplifier LEVEL control
to obtain
a stable display at H
rate.
9.
Check
Delayed Internal
Trigger
Operation
Move the
coaxial cable
connector from P903 to P904
on
the TV sync
separator board and change the Delayed Sweep
Triggering
SOURCE
TV LINE pushbutton. The INT/EXT
pushbutton
is
stored inside the 7B53A/7B53AN OPTION 5
plug-in unit
on the
TV
Sync
Separator board.