background image

7B53A/7B53AN OPTION 5

b.

 

Check

 for a test oscilloscope display of one or more 

negative-going

 pulses

 in a group during the

 vertical sync 

pulse

 interval. The

 number of pulses is dependent on

 the 

television 

standard 

composite video 

signal being applied

 to 

the 

plug-in

 unit. The pulse is very fast; approximately 

0.1 

µs

 

in width

 at the 50%

 amplitude level. Full pulse 

amplitude 

is approximately 2.8 volts.

4. 

Check

 TP920 Composite

 Sync

 Waveform

a.

 

Connect the 10X probe to TP920.

b.

 

Check test oscilloscope display. The 

waveform should 

be

 negative-going composite sync occurring at the H rate; 

amplitude

 should be approximately 180 mV. All video, 

except

 

burst, should 

be clipped off.

5.  Check

 Sync

 Separator Operation

a.

 

Apply

 various television standard composite video 

signals

 (domestic and closed circuit) to the plug-in unit.

NOTE

If desired,

 apply only those composite video signais

 

that

 will be 

used at your facility.

b.  With

 the 10X probe

 connected to TP920, check the 

test 

oscilloscope

 

display for a 

proper waveform as described 

in step 4.

d.

 

Pull

 the DLY TIME knob

 outward and set the DLY 

TIME

 switch to the desired

 position

 (10µs). Rotate the 

DELAY

 TIME MULT control

 to intensify the desired 

portion 

of

 the display. Then, push the DLY TIME knob in 

to view 

the intensified

 portion.

NOTE

In

 this mode of operation the delayed sweep is free

 

running.

7.

  Check

 Discrete Line

 Selector

 Operation

a.

 

Set 

the 7B53A/7B53AN 

OPTION 5

 Main Triggering 

Amplifier

 SOURCE switch to TV LINE. Adjust the LEVEL 

control 

to 

obtain a triggered display.

b.  Rotate the

 DLY'D TRIG control out of the RUNS 

AFTER

 DLY TIME

 position

 and adjust

 this control to 

obtain

 a triggered delayed sweep. Rotate the DELAY TIME 

MULT

 control to

 display the desired portion of the signal.

NOTE

in

 this mode of operation the

 displayed portion of

 

the 

signai will jump

 from line to line as the DELAY

 

TIME

 MUL Tcontrol

 is slowly rotated.

c.

 

Disconnect the 10X

 probe.

6.

 

Check

 Variable Line

 Selector Operation

a.

  Apply

 the

 desired composite video

 signal to the 

vertical 

plug-in 

unit.

b.  Check

 that the 7B53A/7B53AN OPTION 5 Main 

Triggering 

Amplifier SOURCE switch

 is

 set to TV

 FIELD 

and 

the

 

TIME/DIV switch 

is set to 2 ms. Set the TIME/DIV 

VARIABLE 

control 

to

 display

 slightly less than two fields. 

Adjust the

 Main Triggering LEVEL

 control to obtain a 

triggered display.

c.  Set the

 7B53A/7B53AN

 OPTION 5 DLY'D

 TRIG 

control

 to RUNS

 AFTER DLY TIME position. Set the 

Delayed Sweep Triggering controls as follows:

SLOPE
COUPLING 

DC

SOURCE 

TV

 

LINE

8.  Check 

Main

 Triggering Amplifier TV Line Oper­

ation

a.

 

Set

 the TIME/DIV to 10 µs. Check

 that the Main 

Triggering

 

Amplifier SOURCE switch

 is set to TV

 LINE.

b.

 

Adjust the

 Main Triggering Amplifier LEVEL control 

to obtain 

a stable display at H 

rate.

9.

  Check

 Delayed Internal

 Trigger

 Operation

Move the 

coaxial cable 

connector from P903 to P904 

on 

the TV sync 

separator board and change the Delayed Sweep 

Triggering 

SOURCE

 TV LINE pushbutton. The INT/EXT 

pushbutton

 is

 stored inside the 7B53A/7B53AN OPTION 5 

plug-in unit

 

on the

 TV 

Sync

 Separator board.

5

Summary of Contents for 7B53A

Page 1: ...MANUAL SUPPLEMENT TEKTRONIX 7B53A 7B53AN DUAL TIME BASE OPTION 5 INSTRUCTION MANUAL Tektronix Inc 070 1471 00 P O Box 500 Beaverton Oregon 97005 772...

Page 2: ...ect all requests for service or replacement parts to the nearest Tektronix Service Center or Field Office include the type or part number of the product and its serial number Copyright 1972 by Tektron...

Page 3: ...de to P903 or P904 Compatability The TV Sync Separator can be used with 7400 7500 7600 7700 and 7900 Series real time TEKTRONIX oscilloscopes External Graticules External graticules for the main frame...

Page 4: ...less than two fields LINE DISPLAYS 1 Variable Line Selector To obtain Variable Line Selector capabilities the DLY D TRIG control must be set to the RUNS AFTER DLY TIME position Pull the TIME DIV knob...

Page 5: ...e of Q924 The negative pulses will pass through CR918 since during the greater percentage of their duty cycle the line or horizontal sync pulses are positive and will appear to CR918 as positive pulse...

Page 6: ...in a stable display b Adjust the Clipping Level control R914 by rotating the control fully clockwise Then rotate the control counterclockwise until one or more negative going pulses in a group are dis...

Page 7: ...Triggering Amplifier SOURCE switch to TV LINE Adjust the LEVEL control to obtain a triggered display b Rotate the DLY D TRIG control out of the RUNS AFTER DLY TIME position and adjust this control to...

Page 8: ...cer 50 V C932 Add 283 0092 00 0 03 f cer 200 V 80 20 Diodes CR908 Add 152 0075 00 Germanium Tek Spec CR909 Add 152 0141 02 Silicon 1N4152 CR912 Add 152 0141 02 Silicon 1N4152 CR913 Add 152 0141 02 Sil...

Page 9: ...00 150 k 1 4 W 5 R928 Add 315 0822 00 8 2 k 1 4 W 5 R929 Add 315 0123 00 12 k 1 4 W 5 R932 Add 315 0822 00 8 2 k 1 4 W 5 R940 Add 315 0102 00 1 k 1 4 W 5 R941 Add 315 0102 00 1 k 1 4 W 5 S10 Change Sw...