Clear and sanitize procedures
Clear or sanitize
instrument
Use this Secure function to erase con
fi
dential data from your instrument. This
procedure does not erase or change factory calibration constants.
1.
Push the front-panel Utility button to display the Utility menu.
2.
Push the System bezel button.
3.
Push the Secure bezel button. The following message appears on the screen:
“The Secure function will erase all setups and waveforms stored in the
internal memory. Are you sure you want to execute secure function?”
4.
Select OK to execute Secure. Secure does the following:
Erases all waveforms and stored setups in all user memories.
Replaces the current front-panel setup with the default factory setup
values.
Replaces the current GPIB and Ethernet settings with the default factory
settings for models with these ports.
2
AFG2000 Series Declassi
fi