background image

Installation Introductions

 12

016-1674-50 Rackmount Kit

1. Connect the power cord to the rear of the instrument and power source.

Proceed the following procedure as necessary.

2. Connect the AM IN connector on the rear panel of the AFG310/320 through

a 50 

W

 coaxial cable to an appropriate signal source.

3. Connect the SYNC OUT connector on the rear panel of the AFG310/320

through a 50 

W

 coaxial cable to an appropriate equipment which will receive

the SYNC OUT signal.

The cooling air goes inside from the air intakes at top and goes outside from the
rear.
Leave at least 5.1cm(2 inches) free on top when you install the another rack-
mounted instrument above the instrument.  
See Figure 9.

At least the
space of
2 inches(5 cm)
is necessary.

Figure  9: Cooling for Rackmounted instrument(s)

Rear Panel Cabling

Cooling

Summary of Contents for AFG310

Page 1: ...Instructions Manual 016 1674 50 AFG310 AFG320 Rackmount Kit 075 0288 50 www tektronix com 075028850...

Page 2: ...patents issued and pending Information in this publication supercedes that in all previously published material Specifications and price change privileges reserved Tektronix Japan Ltd 5 9 31 Kitashina...

Page 3: ...Description 1 Minimum Tool and Equipment List 1 Kit Parts List 1 Installation Instructions 3 Preparing the Instrument 3 Installing the Instrument into the Rack Adapter 7 Installation into a Equipment...

Page 4: ...ction of the handle retainer cross section 5 Figure 4 Feet and rear cover removal 6 Figure 5 Attaching bracket and panel 8 Figure 6 Attaching bracket to two instruments 9 Figure 7 Attaching panel 10 F...

Page 5: ...ument only in a static free environ ment Observe standard handling precautions for static sensitive devices while installing this kit Always wear a grounded wrist and foot strap while installing this...

Page 6: ...iv 016 1674 50 Rackmount Kit...

Page 7: ...tion 2 Phillips tip Screwdriver Standard tool for M3 and M4 screws Flat bladed screwdriver For removing the plastic cover on the handle bracket Kit Parts List The following table lists the parts inclu...

Page 8: ...Introduction 2 016 1674 50 Rackmount Kit...

Page 9: ...ent to a standard version at a later time 1 Insert the flat bladed screwdriver into the slit between the block and the cover See Figure 1 2 Release the latch of the cover using the flat bladed screwdr...

Page 10: ...ere flat bladed screwdriver may be inserted Figure 1 Location of slit where flat bladed screwdriver may be inserted CAUTION The latch of the cover can be released by lifting it up ward about 5 mm Do n...

Page 11: ...Installation Introductions 016 1674 50 Rackmount Kit 5 M4X10mm M4X10mm Cover Handle Block Screw Cover Block Screw Figure 3 Construction of the handle retainer cross section...

Page 12: ...ips tip screwdriver remove the two screws M4 8 mm securing the rear cover to the instrument See Figure 4 2 Remove the rear cover 3 Install the two screws removed in step 1 to its former screw holes an...

Page 13: ...es include two installation methods H Attaching Bracket H Attaching Panel Procedure for Attaching Bracket to one Instrument 1 Set the instrument with the top down on the work surface and the front fac...

Page 14: ...Installation Introductions 8 016 1674 50 Rackmount Kit M3X6mm M3X6mm Panel Bracket Screw Screw Figure 5 Attaching bracket and panel...

Page 15: ...work surface and the front facing you 2 Put the bracket on the two instruments and align the bracket screw holes with the instruments screw holes See Figure 6 3 Attach the bracket to the instruments w...

Page 16: ...e panel to the bracket with two screws M3 6 mm using a 2 phillips tip screwdriver 3 Set the instrument attached to the bracket with the bottom down on the work surface and the front facing you 4 Attac...

Page 17: ...thods H Installing the adapted instrument into a equipment rack H Rear Panel Cabling 1 Select appropriate mounting holes on the equipment rack frame 2 Attach the bracket to the equipment rack with fou...

Page 18: ...Connect the SYNC OUT connector on the rear panel of the AFG310 320 through a 50 W coaxial cable to an appropriate equipment which will receive the SYNC OUT signal The cooling air goes inside from the...

Page 19: ...016 1674 50 Rackmount Kit 13 Exploded View and Dimensional Drawing 1 2 4 3 3 3 3 Figure 10 Rack adapter exploded view...

Page 20: ...Replaceable Parts 14 016 1674 50 Rackmount Kit 433 427 152 63 152 427 463 6 482 6 21 8 88 44 5 294 350 14 5 30 10 347 80 10 70 88 6 12 ellipse Unit mm Figure 11 Dimensional drawing of rack adapter...

Page 21: ......

Page 22: ......

Reviews: