Basic Test Procedures
CTS850 SDH/PDH Test Set User Manual
3–5
2. Set up the CTS850 as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TEST SETUPS
RECALL
INSTRUMENT
SETUPS
DEFAULT
FACTORY
SETTINGS
Recall Setup
TRANSMIT
TRANSMIT
SETTINGS
Transmit Rate
STM-1E
Payload
Bulk Fill
(Equipped)
Tx/Rx Setup
Coupled
TEST SETUPS
TEST CONTROL
Test Duration
15 Minutes
RESULTS
SDH RESULTS
none
SDh Anomalies
3. Press the START/STOP button.
4. While the test is running, the CTS850 counts errors and errored
seconds and then calculates the error ratio for each error type or
errored seconds.
5. After 15 minutes have elapsed, the test stops automatically and
the measured results are held on the SDH RESULTS page until
another test is started.
6. When the test is complete, you can store the test results to disk or
print out a hardcopy. To save the test results to disk, you must
enter a name for the disk file as follows:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
RESULTS
SAVE RESULTS
Name
EDIT NAME
7. Enter an eight-character name for the file.