Specification—
DM 5010
Table
1-2
(cont)
OHMS
Characteristics
Performance
Requirements
Supplemental
Information
AC VOLTS
—TRUE
RMS
GUARD to LOW
200 V
peak
Rear interface Connector
Pin
28B (HI) to pin
28A (LO)
60 V (d
c plus pk ac)
CREST
FACTOR
4 (subject to maximum
peak input
voltage)
MAXIMUM VOLT
*
Hz PRODUCT
107 V • Hz
TRUE COMMON
MODE
REJECTION (CMR)
Unguarded
With
1
kΩ unbalance in either terminal
Typically
>80 dB from dc to 60 Hz
Guarded
Typically
> 100 dB from dc to 60 Hz
Using NULL
Using NULL on 2k OHM range only.
c
*
Using NULL on 200 OHM range only.
When
the NULL
function is not used, add ±200
milliohms to all readings.
ACCURACY
for the 200 Ω. 2
KΩ,
20
KΩ, 200 KΩ, 2 MΩ, and 20 MΩ
ranges:
Normal
Conversion
Rate
(1.6
readings
per second)
+ 18°C to
+28°C
200
Ω
range
±(0.015% of r 0.015% of
full
scale)
0
2 kΩ to 200 kΩ
ranges
±(0.015% of r
0.01% of full
scale)
0
2 MΩ range
±(0.10%
of
r 0.01% of full
scale)
20 MΩ
range
±(0.15%
of
r
0.005% of full
scale)
0°C
to + 18qC,
+ 28qC
to +50qC
For the typical accuracy for a given
temperature
in this range, add to the
+
18q
C to
+28qC accuracy the following
tolerance,
per degree deviation
from
+
18°C or
+28°C.
200 Ω
range
±(0.06%
of
reading
+ 0.06% of full
scale)
0
2 kΩ to 200 kΩ
ranges
±(0.06%
of
reading
+ 0.035% of full
scale)0
±(0.002% of reading
+ 0.001% of full
scale)/
°C
2 MΩ range
±(0.54% of reading
+
0.035% of full
scale)
±(0.02%
of
reading
+
0.001% of full
scale)/
°C
20 MΩ
range
±(0.9% of
r 0.01% of full scale)
±(0.034% of reading
+ 0.0001% of full
scale)/
°C
1-6
REV
AUG
1984
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......