Performance
Check—DM 5010
Table
5-1
(cont)
Description
Minimum
Requirements
Performance
Check Step
Recommended
Equipment
Resistance
standard
3
200
Ω:
±(0.0038% of rdng
+
0.0038% of range)
2 k-200
kΩ: ±(0.0038% of rdng
+ 0.0025%
of range)
2 MΩ:
±(0.025% of rdng +
0.0025% of
range)
20
MΩ:
±(0.038% of
rdng
+
0.0012% of
range)
5
Electro
Scientific Industries,
Inc.
DB
62 Dekabox
and SR1 10 MΩ
Standard
Resistor
Counter
100
MHz
±0.0016%
4
TEKTRONIX
DC 509
Universal
Counter/Timer
*
Resistor
100
kΩ, ±5%, 1/4 W
3
Tektronix
Part
No. 315-0104-00
Resistor
604
Ω,
±1%, 1/4 W
6
Tektronix
Part
No. 322-0172-00
Controller
GPIB
compatible
9
TEKTRONIX
4050-Series
Controller
or
TEKTRONIX
4041 Controller
Resistance
of Dekabox and interconnect cable must be
known to the accuracy listed
in Table 5-1 for each point checked in the Ohms
Accuracy
check.
Requires a TM 500/5000-Series power module.
Preparation
Make
certain
the
50-60 Hz jumper is positioned to match
the
power module
line frequency. To check the jumper posi
tion,
turn the two
plastic fasteners on the left side cover and
remove
the
cover. Refer to Fig. 10-1
in the pullout pages for
the
jumper location and position.
Replace the side cover.
Install
the
DM 5010 in the
power module, and apply pow
er.
Allow 30 minutes
warm-up time (60 minutes after stor
age in high-humidity environment) before
beginning the
performance check.
Preliminary
Control Settings
DCV
on
NULL
off
LOW FREQ RESPONSE
off
RANGE
STEP (200 mV range)
TRIGGER
MODE
RUN
on
CONVERSION
RATE
FAST off
CALCULATIONS
all off
REAR INPUT
off
WA
RUING
Dangerous voltages
may be encountered in the
fol
lowing
steps. Caution must be exercised. Do
not con
tact
the output connectors
of
the voltage calibrator,
the
input terminals of the DM SOW, or the internal cir
cuitry
of the DM SOW. Set
all voltage
calibrators to a
minimum
output before making
the
necessary
connections.
1.
Check De
Voltage
Accuracy
a. Set the DM 5010
front-panel controls to match the
preliminary
control
settings
listed above.
b. Connect the de voltage calibrator to the
DM 5010
HIGH
and LOW
input connectors through appropriate ca
bles
and adapters.
c.
Set
the
calibrator output to the voltage listed in Table
5-2.
5-2
REV
JAN
1983
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......