Performance
Check—DM
5010
d. CHECK—that the DM 5010 display reads within the
display
limits listed
in the table.
e. Set the CONVERSION RATE (FAST button) as listed
in the next line of the table.
f. CHECK—that the DM 5010
display reads within the
limits listed in the table.
g. Repeat parts c through f for the succeeding
lines of
the table.
h. Set the calibrator to a minimum output voltage.
i. Remove all
connections to
the DM 5010.
Table
5-2
DC VOLTAGE ACCURACY
DM
5010
Range
De Calibrator
Voltage
DM 5010
FAST
Display Limits
Ambient
Temp. Range
+
18 to
+ 28
’C
Ambient
Temp.
Range
0
to +18°C, +28 to
+50
°C
200
mV
0.0
V
off
on
+
0.02 to
—0.02
+ 0.1
to -0.1
+
0.07 to
-0.07
+
0.2 to
-0.2
2 V
0.0 V
on
off
+
.001 to
-.001
+ .0001 to
-.0001
+
.002 to
-.002
+ .0006
to
-.0006
20
V
0.0
V
off
on
+
0.001
to -0.001
+ 0.01
to -0.01
+
0.006 to -0.006
+ 0.02
to
-0.02
200
V
0.0 V
on
off
+ 0.1 to
-0.1
+ 0.01
to -0.01
+
0.2 to
-0.2
+
0.06 to
-0.06
1000
V
0.0
V
off
on
+0.1
to -0.1
+
1. to
-1.
+
0.4 to -0.4
+
2. to —2.
200 mV
190
mV
off
on
189.95
to 190.05
189.8 to 190.2
189.82 to
190.18
189.6 to 190.4
2
V
1.9 V
on
off
1.898
to 1.902
1.8996 to
1.9004
1.896 to 1.904
1.8983 to
1.9017
20
V
19.
V
off
on
18.996
to
19.004
18.98 to 19.02
18.983 to
19.017
18.96 to 19.04
200
V
190.
V
on
off
189.8 to
190.2
189.96 to 190.04
189.6 to 190.4
189.83 to
190.17
1000
V
950. V
off
on
949.7 to 950.3
948. to
952.
949.0 to
951.0
947. to 953.
Set
the
calibrator output to a minimum level.
200
mV
-190
mV
Off
on
-189.95 to
-190.05
-189.8 to
-190.2
-189.82
to
-190.18
-189.6
to
-190.4
2
V
-1.9 V
on
off
-1.898
to
-1.902
-1.8996 to
-1.9004
-1.896
to
-1.904
-1.8983 to -1.9017
20
V
-19. V
off
on
-18.996 to
-19.004
-18.98 to
-19.02
-18.983 to
-19.017
-18.96
to
-19.04
200
V
-190. V
on
off
-189.8 to
-190.2
-189.96 to
-190.04
-189,6
to
-190.4
-189.83 to
-190.17
1000
V
-950. V
off
on
-949.7
to
-950.3
-948. to
-952.
-949.0
to
-951.0
-947. to
-953.
REV
AUG
1984
5-3
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......