Section
6
—DM 5010
ADJUSTMENT
PROCEDURE
Introduction
This procedure should
be performed if the instrument
fails
to meet
the
performance requirements of the electrical
characteristics
listed
in the Specification section of this man
ual.
To ensure continued instrument accuracy, adjustment
should
be performed every 1000 hours of operation or every
six
months
if
used infrequently. Adjustment is also recom
mended
following
instrument
repair or modification. Adjust
ments
must be made at
an ambient temperature between
+ 21 °C to
+
25’C.
Allow thirty minutes
warm-up time
be
fore beginning adjustments
(sixty minutes after exposure to
or storage
in
high humidity environment).
The recommended interval for battery replacement is ap
proximately
every two years. Performance of the Adjust
ment
Procedure is necessary after battery replacement to
restore
the calibration
factors to memory. See the
Mainte
nance
section of this manual for battery replacement
information.
Services Available
Tektronix, Inc.
provides complete instrument repair and
adjustment at
local field service centers and at the factory
service center.
Contact your local Tektronix Field Office or
representative for further
information.
Test
Equipment Required
The
test equipment
listed in Table 6-1, or equivalent, is
recommended
for
adjustment of
the DM5010.
Specifica
tions given for the
test equipment are the minimum neces
sary for accurate instrument adjustment. All
test equipment
is
assumed to be correctly calibrated
and operating within
specification.
Table
6-1
TEST
EQUIPMENT LIST
Description
Minimum
Requirements
Performance
Check Step
n
--------------—--------------------------------------
Recommended
Equipment
TM
5000-Series
Power
Module
all
TEKTRONIX TM
5003,
Opt.
02
DC
voltage calibrator
200
mV: ±(0.0038% of rdng
+
0.0025% of range)
2V-200V: ±(0.0038% of
rdng
+
0.0013% of range)
1000 V:
±(0.005% of rdng +
0.0025% of range)
2
Fluke
335D De
Voltage Standard
Ac
voltage calibrator
200 mV through 200 V:
10-20 Hz: ±(0.20% of rdng
+
0.08%
of
range)
20-100 Hz:
±(0.20% of rdng
+
0.05%
of
range)
100 Hz-20 kHz: ±(0.05% of
rdng
+ 0.05% of range)
20-100
kHz:
±(0.25% of rdng +
0.12% of range)
6, 7
Fluke 5200A
Ac Calibrator, and
Fluke
5215A Power Amplifier
6-1
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......