Operating instructions
—DM 5010
FRONT
PANEL
CONTROLS
AND
CONNECTORS
General Information
The
seventeen front-panel
push buttons listed below
en
able their functions and illuminate when
*
The push
buttons
in the left column are self-cancelling; that is, only
one
push
button is active
(illuminated) at a time. The push
buttons
in the right column
illuminate when activated, and
remain
illuminated and
active
until pressed again.
DCV
NULL
OHMS
LOW FREQ RESPONSE
DIODE TEST
AUTO
ACV
RUN
ACV
+
DCV
TRIGGERED
FAST
AVERAGE
X-B
A
dBm
dBr
COMPARE
REAR
INPUT
The remaining
twenty-three
are non-illuminating push
buttons that activate
their associated functions. Refer to
Fig.
2-2.
Display
Window
The left
side of
the window
displays
measurements
and
calculation
results using signed
4
1/2
digit LEDs. Zeros lead
ing the decimal
are suppressed. A flashing display
indicates
over-range
when the voltage measurement functions are ac
tive; OC is displayed
for OHMS and DIODE TEST.
The
center area of the window indicates the range multi
plier
for
the displayed
reading. Illuminated
LED associated
with
the
words
MILLI, KILO,
and MEGA on the window indi
cate
when the
displayed measurement is in milli, kilo, or
mega units.
The decimal point is fixed for each function
range.
The
multiplier LED and decimal point location indi
cate the
range for both AUTO and manual (STEP) methods
of
range selection. See Fig. 2-2.
The right area of the display window indicates the oper
ating state
of the instrument,
as follows:
REMOTE
and
ADDRESSED illuminate only when the in
strument is operating under remote
program control via
the GPIB.
ERROR illuminates
when an internal
error, seif test error,
or
operating
error occurs. The left area of the window
also displays front panel error codes indicating the type
of
error. See
Table 2-2
in Operators Familiarization in this
section
of the
manual.
Function
Illuminated
Multiplier
Selecte d Range
DCV
• MILLI
200 mV
2
V
20 V
200
V
1000
V
ACV
and
ACV
+
DCV
•MILLI
200
mV
2
V
20
V
200
V
700
V
OHMS
200 Ω
•KILO
2
kΩ
20
kΩ
200
kΩ
•
MEGA
2
MΩ
20
MΩ
DIODE
TEST
Uses
2 V
dc
range.
Blank
when
FAST CONVERSION
RATE
is enabled. 2994.02
Overrange
indication:
Display
flashes for voltage functions.
Displays "OC” for OHMS and
DIODE TEST functions.
Fig. 2-2. DM 5010 front panel range indication.
2-2
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......