Programming
—DM 5010
SET?
SOURCE
Type:
Type:
Query
Setting
or
query
Query
syntax:
Setting syntax:
SET?
Query
response example
(power up settings):
SOURCE
FRONT
SOURCE REAR
SOUR
FRONT
SOUR
REAR
DCV
— 1.E+3;AVE 2; RATIO 1. O.;DBR 1.;LIMITS 0.,
O.jCALC OFF;NULL
0.;
DIGIT 4.5;LFR OFF;MODE
RUN;SOURCE
FRONT;DT OFF;MONITOR OFF;OPC
OFF;
OVER
OFF;USER
OFF;RQS ON;
Query
syntax:
SOURCE?
SOUR?
Discussion:
This
command returns the present settings of all instru
ment functions.
The
longest response is 225 characters.
Query
response:
SOURCE
FRONT;
SOURCE REAR;
Discussion:
SOURCE
FRONT selects the front panel input for mea
surement;
SOURCE REAR
selects the rear interface input
connectors for
measurement.
Do
not
switch from
front panel to rear interface input
while
over 500 V peak is
applied to the front pane!
input connectors.
Instrument damage and erratic op
eration
may result.
SET?
SOURCE
3-19
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......