TB 9-6625-1957-35
7
Figure 1. Tektronix, Type DM 501 - left side view.
9. Resistance
a. Performance Check
(1) Connect calibrator
OUTPUT
to TI
INPUT HI
and
LO
terminals.
(2) Set
RANGE/FUNCTION
switch to
200K OHMS
.
(3) Set calibrator for a 190 k
Ω
output. Adjust calibrator's output adjustment
control for a calibrator control display reading equal to the TI's indication. The calibrator's
control display
ERROR
indication will be
≤
±0.31 percent; if not, perform
b
below.
(4) Repeat technique of (2) and (3) above using values listed in table 4. Calibrator's
ERROR
display will be within the limits specified.
b. Adjustments
. Adjust R35
Ω
CAL (fig. 1) for a T1 indication equal to calibrator's
displayed value for 190 k
Ω
(R).