TB 9-6625-1957-35
4
Table
1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Tektronix, Type DM 501A - Continued
Resistance
Range: 0 to 20 M
Ω
(in 5 ranges)
Accuracy:
±(% of r % of FS)
Range:
200
Ω
through 200 k
Ω
(LO
Ω
) ................... 0.15 + .015
2 k
Ω
through 2000 k
Ω
(HI
Ω
)................... 0.15 + .015
2000 k
Ω
(LO
Ω
) ........................................ 0.3 + .015
20 M
Ω
(HI
Ω
) .......................................... 0.5 + .015
Dc current
Range: 0 to 2000 mA (in 5 ranges)
Accuracy:
±
(0.2% of r .015% of FS)
Dc High Voltage Probes (All Models)
Dc voltage
Range: 1 to 40 kV
2
Accuracy:
4.00% at 1 and 40 kV
3.0% at 10 kV
2.0% at 20 and 30 kV
±1.0% at 25
1
Accuracy applies to TI display value. To obtain correct dB reading, algebraically add the range selected
to the display reading. Dynamic range is -60 to +56.2 dBrn.
2
Verified at 1000 V only.
SECTION II
EQUIPMENT REQUIREMENTS
Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
5. Accessories Required.
The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph
above, and are not listed in this
calibration procedure.
Table 2. Minimum Specifications of Equipment Required
Common name
Minimum use specifications
Manufacturer and model
(part number)
CALIBRATOR
Dc voltage:
Range: .19 to 1000 V
Accuracy: ±0.14%
Ac voltage:
Range: 100 mV to 500 V
Frequency: 20 Hz to 20 kHz
Accuracy:
±
(%)
1.0 to 10 kHz ...................... 0.16
20 Hz and 20 kHz ............... 0.26
John Fluke, Model 5700A/CT (p/o
MIS-35947); w/power amplifier
John Fluke, Model 5215A/CT
(5215A/CT); w/transconductance
amplifier, John Fluke, Model
5220A/CT (5220A/CT); w/ac
divider, John Fluke, Model 7405A-
4207 (7405A-4207)
Summary of Contents for DM501A
Page 1: ......
Page 2: ......
Page 6: ...OM 501A OM SOiA Oigita Multlmeter xiv AUG 1980...
Page 13: ......
Page 18: ...Verso Filler Page z z...
Page 21: ...Operating Instructions OM SOlA FIg 2 2 Controls and connectors REV A AUG 1980 English 2 3...
Page 24: ...Verso Filler Page z z...
Page 25: ...OM 501A Japanese 2 3...
Page 26: ...Verso Filler Page z z...
Page 48: ...Verso Filler Page z z...