Connecting the probe to your circuit
Attach the probe to the circuit using the connectors and adapters shown on the back of these instructions. Select the best
method for your needs, and then proceed to Setting up the Probe.
Setting up the probe
To set and view the digital channel parameters, do the following:
On MSO/DPO5000B Series instruments, select Vertical>Digital Setup or push the
D15-D0
button.
On MSO/DPO4000B Series instruments, push the
D15–D0
button.
The parameters listed below can be set on each digital channel:
Threshold voltage and vertical position (the default threshold setting is 1.4 V)
Signal height and position (set once for all 16 channels)
Channel label
To set and view bus characteristics, do the following:
On MSO/DPO5000B Series instruments, use the controls in the Bus Setup menu.
On MSO/DPO4000B Series instruments, push the
B1
to
B4
buttons.
The setup screens allow you to set and view bus characteristics such as:
Clock type
Bus type (Serial or Parallel)
Bus width
Display format (Hex, Binary, or ASCII symbols)
Parallel bus setup information is resident on some oscilloscope models. However, for other buses such as SPI and I2C, you
must have the appropriate option. See your oscilloscope manual or product data sheet for nomenclature and ordering details.
Functional check
Logic activity immediately displays on all connected, active channels. If you do not see an active signal:
1.
Push the
Trigger
button.
2.
Select Edge for trigger type.
3.
Select the channel that you are setting up as the source.
4.
On MSO/DPO4000B Series instruments, push the
Autoset
button.
If you do not see an active signal, try another probe channel (or analog probe) to verify circuit activity at the test point.