Install your instrument
7.
Maximum input voltage, DPO7000C
models:
50
Ω
1 M
Ω
5 V
RMS
, with no impulse allowed, with peaks
≤
±24 V.
150 V, derate at 20 dB/decade to 9 V
RMS
above 200 kHz. The maximum
input voltage at the BNC, between center conductor and ground is
400 V peak. The RMS voltage is limited to <150 V for arbitrary waveshapes
including DC. The maximum pulse width for impulses with peaks over
150 V is 50
μ
s. Example: At 0 V to 400 V peak, rectangular wave, the duty
factor is 14%. The maximum transient withstand voltage is ±800 V peak.
Maximum input voltage, 4 GHz to 20 GHz
models:
50
Ω
<1 V
RMS
for <1 V/FS settings and < 5.5 V
RMS
for
≥
1 V/FS settings.
Maximum input voltage, > 20 GHz models:
50
Ω
≤
1.2 V/FS settings:
±1.5 V relative to the termination bias (30 mA maximum).
±5 V absolute maximum input.
>1.2 V/FS settings:
10 V at 25 °C derated to 8.4 V at 45 °C (limited by attenuator).
Maximum nondestructive input voltage to
logic probes, MSO70000C/DX Series:
±15 V
CAUTION.
For proper cooling, keep the bottom and sides of the instrument clear of obstructions.
MSO5000B and DPO5000B
1.
Place the instrument on a cart or bench.
The instrument should rest on its bottom
or rear feet. An optional rack mounting
kit is available. Observe the following
clearance requirements and dimensions:
Rear:
2 in (50.8 mm)
Left side:
2 in (50.8 mm)
2.
Width:
17.3 inches (439 mm)
3.
Height:
9.2 inches (233 mm) including feet
10.7 inches (272 mm) including vertical handle and feet
4.
Before operating the instrument, verify
the ambient temperature:
5 °C to +50 °C (+41 °F to +131 °F)
4