Install your instrument
Operating requirements
MSO/DPO70000DX, MSO/DPO70000C, and DPO7000C
1.
Place the instrument on a cart or bench.
The instrument should rest on its bottom
or rear feet. An optional rack mounting
kit is available. Observe the following
clearance requirements and dimensions:
DPO7000C Models:
MSO/DPO70000C/DX Models:
Top:
0 in (0 mm)
0 in (0 mm)
Left and right side:
0 in (0 mm) on right side
3 in (76 mm) on left side
3 in (76 mm)
Bottom:
0 in (0 mm) standing on feet,
fl
ip
stands down
0 in (0 mm) standing on feet,
fl
ip
stands down
Rear:
0 in (0 mm) on rear feet
0 in (0 mm) on rear feet
2.
Width:
17.96 inches (456 mm)
17.75 inches (451 mm)
3.
Height:
10.9 inches (277 mm)
11.48 inches (292 mm)
4.
Before operating the instrument, verify
the ambient temperature:
5 °C to +45 °C (+41 °F to +113 °F)
5 °C to +45 °C (+41 °F to +113 °F)
5.
Verify the operating humidity:
8% to 80% relative humidity with
a maximum wet-bulb temperature
of +29 °C (+84 °F) at or below
+45 °C (+113 °F), noncondensing
Upper limit derated to 30% relative
humidity at +45 °C (+113 °F)
8% to 80% relative humidity at up
to +32 °C (+90 °F)
5% to 45% relative humidity above
+32 °C (+90 °F) up to +45 °C
(+113 °F), noncondensing, and
is limited by a maximum wet-bulb
temperature of +29.4 °C (+85 °F)
(derate relative humidity to 32% at
+45 °C (+113 °F))
6.
Verify the operating altitude:
DPO7000C models: 3,000 m
(9,843 feet)
MSO/DPO70000C/DX models:
3,000 m (9,843 feet), derate
maximum operating temperature
by 1 °C per 300 meters
(984.25 feet) above 1500 meters
(4921.25 feet) altitude.
MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series User Manual
3