background image

Model 2290-10-SHVUC 

SHV Male to Unterminated Cable

 

 

 

 

077084500     

*077084500*

 

Keithley Instruments, Inc.

 

28775 Aurora Road 
Cleveland, Ohio 44139 
1-888-KEITHLEY 

http://www.keithley.com

 

Description 

The Model 2290-10 can contain hazardous voltages. Be certain that the high voltage is 
completely discharged before removing or connecting the high-voltage cable. High-voltage 
cables can store charge if they are disconnected from the supply while high voltage is turned 
on and can cause personal injury or death if not handled properly. 
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the 
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth 
ground using #18AWG or larger wire. 

The high-voltage 10 kV SHV unterminated cable connects to the Model 2290-10 SHVBH female bulkhead 
connector (refer to document number 077084700 for more information). 

High-voltage installation: 

1.  Turn off power and allow all voltage to discharge. 

2.  Install the Model 2290-5-SHVUC cable to the Model 2290-10 SHVBH female bulkhead connector (see next 

figure). 

Figure 1: Model 2290-10 SHVUC male to unterminated cable 

 

 
 