Connect
Install the instrument
You can use a Series 3700A instrument on a bench or in
a rack. Please see the instructions that came with your
rack‑mount kit if you are installing a Series 3700A instrument
in a rack.
Note that the air intakes for the fan are on the side panel of
the instrument. The space around these areas should be free
from obstruction to ensure proper fan operation.
Connect line power
Series 3700A instruments operate from a line voltage of
100 V to 240 V at a frequency of 50 Hz or 60 Hz. Line
voltage is automatically sensed (there are no switches
to set). Make sure the operating voltage in your area is
compatible.
CAUTION
Operating the instrument on an incorrect line voltage
may cause damage to the instrument, possibly voiding
the warranty.
WARNING
The power cord supplied with the Series 3700A contains
a separate protective earth (safety ground) wire for use
with grounded outlets. When proper connections are
made, the instrument chassis is connected to power‑line
ground through the ground wire in the power cord. In
the event of a failure, not using a properly grounded
protective earth and grounded outlet may result in
personal injury or death due to electric shock.
WARNING
Do not replace detachable mains supply cords with
inadequately rated cords. Failure to use properly rated
cords may result in personal injury or death due to
electric shock.