background image

 

PA-289 Rev. D April 2021 

 

The 

CAUTION

 heading in the user documentation explains hazards that could damage the instrument. Such damage may 

invalidate the warranty. 

The 

CAUTION

 heading with the 

  symbol in the user documentation explains hazards that could result in moderate or minor 

injury or damage the instrument. Always read the associated information very carefully before performing the indicated 

procedure. Damage to the instrument may invalidate the warranty.   
Instrumentation and accessories shall not be connected to humans. 
Before performing any maintenance, disconnect the line cord and all test cables. 
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power 

transformer, test leads, and input jacks — must be purchased from Keithley. Standard fuses with applicable national safety 

approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may 

only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other 

suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through 

Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement 

component, call a Keithley office for information. 
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following 

environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2. 
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the 

instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that 

consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never 

require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board 

should be returned to the factory for proper cleaning/servicing. 
Safety precaution revision as of June 2017. 

 

Summary of Contents for KEITHLEY 4288-1

Page 1: ...art number 1 Front rack ear 4288 1 302 1 Dress panel 4288 1 301 4 Cage nut FA 274 4 8 32 3 8 in Phillips pan head screw 8 32 3 8PPH 4 10 32 5 8 in Phillips truss head screw 10 32 5 8PHTRSH Tools required Medium 2 Phillips screwdriver 3 8 inch wrench Flat bladed screwdriver or cage nut installation tool Installation overview Review the Installation precautions on page 2 Select a location in the rac...

Page 2: ...nt damaging heat build up and ensure specified performance make sure there is adequate ventilation and airflow around the instrument to ensure proper cooling Do not cover the ventilation holes on the sides or back of the instrument The rear vents and both side vents must be unobstructed to properly dissipate heat Mount precision equipment as low as possible in the rack Operating temperatures are c...

Page 3: ... 1 away from the case 3 Remove the screw that secures each mounting ear 3 4 Pull down and out to remove each ear Figure 2 Mounting ear removal 5 From the bottom of the instrument pull out the rubber feet and remove the screws 4 see the following figure Figure 3 Instrument rubber feet removal 6 Store the handle rubber feet and hardware for future use ...

Page 4: ...o the side without the instrument To install the mounting hardware on the instrument 1 Select the side that will be closest to the rack rail see note above 2 Secure the front rack ear 4288 1 302 to one side of the instrument using two Phillips pan head screws 8 32 3 8 PPH Figure 4 Attach rack ear 3 Secure the dress panel 4288 1 301 to the other side of the instrument using two Phillips pan head sc...

Page 5: ... rack Install the cage nuts FA 274 into the front rack holes for the front rack ear 4288 1 302 and the dress panel 4288 1 301 Cage nuts FA 274 for the front rack ear 4288 1 302 and dress panel 4288 1 301 should be installed into the rails at the same hole spacing vertical height Figure 5 Rack mount rails ...

Page 6: ...e to disconnect all power when needed may expose personnel to hazardous voltages which if contacted could cause personal injury or death To complete this procedure two people are recommended 1 Slide the instrument into the rack by aligning the front rack ear 4288 1 302 and the dress panel 4288 1 301 with the cage nuts FA 274 in the rack rails 2 While supporting the instrument secure it to the fron...

Page 7: ...ts will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a shock hazard is present Lethal voltage may be present on cable connector jacks or test fixtures The American National Standards Institute ANSI states that a shock hazard exists when voltage levels gr...

Page 8: ...r components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithley to maintain accuracy and functionality of the product If you are unsure about the applicability of a replacement component call a Keithley office for information Unless otherwise noted in product s...

Reviews: