background image

 

PA-909 Rev. J April 2021 

11 

 

 

 

Safety precautions 

The following safety precautions should be observed before using this product and any associated instrumentation. Although 

some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous 

conditions may be present. 
This product is intended for use by personnel who recognize shock hazards and are familiar with the safety precautions required 

to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the 

product. Refer to the user documentation for complete product specifications. 
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. 

 

The types of product users are: 

Responsible body

 is the individual or group responsible for the use and maintenance of equipment, for ensuring that the 

equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained. 

Operators

 use the product for its intended function. They must be trained in electrical safety procedures and proper use of the 

instrument. They must be protected from electric shock and contact with hazardous live circuits. 

Maintenance personnel

 perform routine procedures on the product to keep it operating properly, for example, setting the line 

voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures 

explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel. 

Service personnel

 are trained to work on live circuits, perform safe installations, and repair products. Only properly trained 

service personnel may perform installation and service procedures. 

 

Keithley products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low 

transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient 

overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient 

overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to 

mains. These instruments will be marked as category II or higher. 
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains. 
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test 

fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 

30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any 

unknown circuit before measuring. 
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators 

are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential 

human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If 

the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed. 
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under 

test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting 

cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers. 

 

Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) 

ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the 

voltage being measured. 
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or 

accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be 

impaired. 
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the 

specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards. 
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground) 

connections. 
The 

WARNING

 heading in the user documentation explains hazards that might result in personal injury or death. Always read 

the associated information very carefully before performing the indicated procedure. 

Summary of Contents for Keithley 4299-2 Dual-Unit

Page 1: ...ts included in the kit Quantity Description Keithley part number 2 Hinge 2602 328 2 Front rack ear 2602 355 2 Rear rack mount bracket 2602 334 2 Rack mount extension arm 2602 335 1 Rear bracket 2304 327 1 Quick release pin FA 352 8 Cage nut FA 274 2 6 Lock washer 6LKWA 2 Washer WA 81 4 8 32 3 8 in Phillips pan head screw 8 32 3 8PPH 4 8 32 3 8 in Phillips flat head screw 8 32 3 8PFH 4 10 32 3 8 in...

Page 2: ...nd unplug all cables before installing an instrument in a rack Failure to remove power before installation may cause personal injury or death from electrical shock Heat sources should be mounted away from the instrument cabling and accessories with sufficient space provided between the instrument and heat source for airflow and cooling To prevent damaging heat build up and ensure specified perform...

Page 3: ...rientation arrows on the mounting ears as shown in the following figure Figure 1 Rack mount kit handle removal 2 After you align the arrows 2 pull the ends of the handle 1 away from the case 3 Remove the screw that secures each mounting ear 3 4 Pull down and out to remove each ear Figure 2 Mounting ear removal 5 From the bottom of the instrument pull out the rubber feet and remove the screws 4 see...

Page 4: ... by side in the same position that they will be installed in the rack 2 Secure one rear rack mount bracket 2602 334 to each instrument on the selected side using two Phillips pan head SEMS screws 10 32 3 8 PPHSEM If the two instruments are not the same length do not use one of the rear rack mount brackets Refer to Install the instruments on page 9 To install two instruments that are of unequal len...

Page 5: ...lation Instructions PA 909 Rev J April 2021 5 3 Attach the front rack ear 2602 355 to the outside of each instrument using two Phillips pan head screws 8 32 3 8 PPH on the same side as the rear rack mount bracket 2602 334 Figure 5 Front rack ears ...

Page 6: ...2602 328 to both instruments using two Phillips flat head screws 8 32 3 8 PFH on each make sure the knuckle of the hinge is facing the front of the instrument as shown in the following figure 5 Insert the quick release pin FA 352 in the hinge knuckle to assemble the hinge Figure 6 Secure hinges to instrument ...

Page 7: ...H washers WA 81 and lock washers 6LKWA Figure 7 Install rear bracket Prepare the rack To prepare the rack for installation 1 Install the cage nuts FA 274 into the rack holes to be aligned with the holes in the rack mount extension arms 2602 335 The rack holes selected determine the placement of the instrument in the rack see the following figure 2 Install one rack mount extension arm onto one of t...

Page 8: ... one of the rack mount extension arms Refer to Install the instruments on page 9 To install two instruments that are of unequal length for more detail Figure 8 Rack mount extension arms 3 Repeat the procedure on the other side 4 Install cage nuts into the front rack rail holes aligned with the rear cage nuts 5 Save all unused parts for future use ...

Page 9: ...ll power when needed may expose personnel to hazardous voltages which if contacted could cause personal injury or death To complete this procedure two people are recommended To install the instruments 1 Slide the instruments into the rack by aligning both rear rack mount brackets 2602 334 with the rack mount extension arms as shown in the following figure 2 While supporting the instruments secure ...

Page 10: ...o the rack by aligning the instrument rear rack mount bracket 2602 334 and the rack mount extension arm 2602 335 also align the front rack rail for the other instrument as shown in the following figure 2 While supporting the instruments secure them to the front rack rails with four Phillips truss head screws 10 32 5 8 PHTRSH Figure 10 Securing different length rack rails ...

Page 11: ...nts will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a shock hazard is present Lethal voltage may be present on cable connector jacks or test fixtures The American National Standards Institute ANSI states that a shock hazard exists when voltage levels g...

Page 12: ...r components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithley to maintain accuracy and functionality of the product If you are unsure about the applicability of a replacement component call a Keithley office for information Unless otherwise noted in product s...

Reviews: