Kelvin Keithley Triaxial Connector Card 

 

 

 

 

071321300 / October 2014      

*P071321300*

 

 

Keithley Instruments

 

28775 Aurora Road 
Cleveland, Ohio 44139 
1-800-935-5595 

http://www.keithley.com

 

Description 

The Model 8020-KHV Keithley HV Connector Card provides high-voltage triaxial connections for sense and 
force connections on the Model 8020 High Power Interface Panel. Use this connector card to connect the 
Model 8020 to a device under test using the Keithley high-voltage triaxial connector. The connector card can be 
installed in any Model 8020 channel. 

Figure 1: Model 8020-KHV, front view 

 

Figure 2: Model 8020-KHV, side view 

 

 

 

Summary of Contents for Keithley 8020-KHV

Page 1: ... 8020 KHV Keithley HV Connector Card provides high voltage triaxial connections for sense and force connections on the Model 8020 High Power Interface Panel Use this connector card to connect the Model 8020 to a device under test using the Keithley high voltage triaxial connector The connector card can be installed in any Model 8020 channel Figure 1 Model 8020 KHV front view Figure 2 Model 8020 KH...

Page 2: ...del 8020 Voltage 3070 V maximum Connector current fuse installed 1 A maximum 10 A pulsed 1 ms at 2 2 percent duty cycle Connector current fuse removed 1 ADC no pulsing Connector types 3 lug triaxial Fuse specifications 10 A 32 V MINI blade style Figure 3 Connector conductors when installed in the HV and 2600 4200 channels of the Model 8020 Figure 4 Connector conductors when installed in the Model ...

Page 3: ...071321300 October 2014 3 Dimensions Figure 5 Model 8020 KHV dimensions Installation See the Model 8020 User s Manual for installation information available for download on the Keithley Instruments website Support page ...

Page 4: ...overcurrent protection This transient overcurrent protection is useful for protecting probes and other sensitive interface hardware at the device under test For more information see the Model 8020 High Power Interface Panel User s Manual Transient overcurrent protection is only available on the 200 V channels within the Model 8020 When using a connector card in the HV channel be sure that the 10 A...

Page 5: ... be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a shock hazard is present Lethal voltage may be present on cable connector jacks or test fixtures The American National Standards Institute ANSI states that a shoc...

Page 6: ...ly disposed of according to federal state and local laws The WARNING heading in the user documentation explains dangers that might result in personal injury or death Always read the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that could damage the instrument Such damage may invalidate the warranty In...

Reviews: