DAQ6510 Data Acquisition and Multimeter System Calibration and Adjustment Manual
Section 2: Performance verification
DAQ6510-905-01 Rev. D June 2022
2-5
Performing the verification test procedures
The following topics provide a summary of calibration verification test procedures and items to
consider before performing any calibration verification test.
Test summary
Front-panel tests:
•
(on page 2-6)
•
(on page 2-9)
•
(on page 2-14)
•
(on page 2-17)
•
Simulated thermocouple type J temperature verification
(on page 2-18)
•
Simulated RTD temperature verification
(on page 2-20)
•
(on page 2-24)
•
(on page 2-29)
•
(on page 2-33)
•
(on page 2-36)
•
(on page 2-39)
•
Verifying zero values using a 4-wire short
(on page 2-41)
Test considerations
When performing the calibration verification procedures:
•
Be sure to restore factory front-panel defaults. From the front panel, select the
MENU
key, select
Info/Manage
, and select
System Reset
.
•
Make sure that the test equipment is warmed up for the time recommended by the manufacturer
and is connected to the DAQ6510 input/output terminals.
•
Make sure that the correct DAQ6510 terminals are selected with the TERMINALS FRONT/REAR
switch.
•
Make sure the test equipment is set up for the proper function and range.
•
Do not connect test equipment to the DAQ6510 through a scanner, multiplexer, or other switching
equipment.